FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 8903175 · Received August 16, 2019

Report

Report Number
9617229-2019-10493
Event Type
Injury
Date Received
August 16, 2019
Date of Event
July 8, 2019
Report Date
September 20, 2019
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000391
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASE FLAT, WEAR ABRASION AND OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: THE UNIT DOES NOT LEAK. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO WERE OBSERVED OPENINGS ON THE DEVICE OR ISSUES RELATED WITH THE COMPLAINT.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED A LEFT-SIDE DEFLATION. DEVICE WAS EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. REASON FOR REOPERATION: DEFLATION. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A LEFT-SIDE DEFLATION. DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
695835 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 3133424 10888628000391

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention