FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 8899445 · Received August 15, 2019

Report

Report Number
3013756811-2019-48029
Event Type
Malfunction
Date Received
August 15, 2019
Date of Event
July 23, 2019
Report Date
August 15, 2019
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE INTERMITTENT INACCURACIES BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) READING AND THE BLOOD GLUCOSE (BG) METER READING. REPORTEDLY, THE CGM BG READINGS WERE 70-190 MG/DL, AND THE METER BG READINGS WERE 120-233 MG/DL. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. REPORTEDLY, THE CUSTOMER PERFORMED A CALIBRATION AND CGM READINGS BECAME ACCURATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693231 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-02

Patients

Seq Age Sex Outcome Treatment
1 31 YR