FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY

MDR report key: 8898902 · Received August 15, 2019

Report

Report Number
9610847-2019-00516
Event Type
Malfunction
Date Received
August 15, 2019
Date of Event
July 17, 2019
Report Date
September 16, 2019
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
30382903096801
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9091843. OUR RECORDS SHOW THAT THIS IS THE SECOND INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. COMPOSITION TESTING OF THE FOREIGN MATERIAL DETERMINED THAT THE SAMPLE TRAYS CONTAINED CARDBOARD DEBRIS, WHICH ORIGINATES FROM THE SYRINGE COMPONENT BOXES. THE COMPONENTS ARE DOUBLE BAGGED IN ORDER TO PREVENT THE DEBRIS FROM CLINGING TO THE BAG WHILE THE SYRINGES ARE LOADED INTO THE MACHINE HOPPER. IN THE EVENT MANUFACTURING PERSONNEL FAIL TO FOLLOW ESTABLISHED MANUFACTURING GUIDELINES, CARDBOARD CAN BE TRANSFERRED INTO THE FINISHED GOODS.

Description of Event or Problem · 0

IT WAS REPORTED THAT 3 BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAYS WERE FOUND BEFORE USE WITH CARDBOARD IN THEM. THIS COMPLAINT WAS CREATED TO CAPTURE THE 3RD OF 4 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "3 TRAYS WERE DISCOVERED TO HAVE CARDBOARD IN THEM."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 3 BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAYS WERE FOUND BEFORE USE WITH CARDBOARD IN THEM. THIS COMPLAINT WAS CREATED TO CAPTURE THE 3RD OF 4 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "3 TRAYS WERE DISCOVERED TO HAVE CARDBOARD IN THEM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690190 BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 9091843 30382903096801

Patients

Seq Age Sex Outcome Treatment
1 Other