FDA Adverse Event Malfunction Summary report: N

PRN ADAPTER

MDR report key: 8897568 · Received August 15, 2019

Report

Report Number
3006948883-2019-00654
Event Type
Malfunction
Date Received
August 15, 2019
Date of Event
July 22, 2019
Report Date
September 9, 2019
Manufacturer
BD (SUZHOU)
Product Code
FPA
UDI-DI
00382903886388
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8233030. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ALTHOUGH PHOTOS WERE SUBMITTED FOR EVALUATION, THEY DID NOT DISPLAY THE FAILURE MODE CLEARLY ENOUGH TO IDENTIFY THE ROOT CAUSE. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION, AND WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE SECTION H.10

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND BEFORE USE WITH A PRN ADAPTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "IT¿S NOTICED THAT YELLOW FOREIGN MATTER WAS EMBEDDED IN MALE LUER CONNECTION BEFORE UNWRAPPED THE PACKAGE DURING PRIMING."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND BEFORE USE WITH A PRN ADAPTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "IT¿S NOTICED THAT YELLOW FOREIGN MATTER WAS EMBEDDED IN MALE LUER CONNECTION BEFORE UNWRAPPED THE PACKAGE DURING PRIMING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693662 PRN ADAPTER ADAPTER FPA BD (SUZHOU) 8233030 00382903886388

Patients

Seq Age Sex Outcome Treatment
1 Other