FDA Adverse Event Malfunction Summary report: N

BD SYRINGE S2 5ML

MDR report key: 8896541 · Received August 15, 2019

Report

Report Number
3002682307-2019-00451
Event Type
Malfunction
Date Received
August 15, 2019
Date of Event
July 22, 2019
Report Date
August 21, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAS BEEN PROVIDED WITH A PHOTO FOR CATALOG 309050 LOT 1903206 TO INVESTIGATE FOR THIS RECORD. THE FOREIGN MATTER WAS IDENTIFIED AS PRINTING FOIL PARTICLES IN THE FLUID PATH. AS A RESULT, BD WAS ABLE TO VERIFY THE REPORTED ISSUE. THE PROCESS USED TO PRINT THE SCALE IN BD EMERALD SYRINGES IS CALLED "HOT STAMPING" PROCESS, IN WHICH, THROUGH A HEATED METAL STAMP, THE SCALE IS TRANSFERRED FROM THE PRINTING FOIL TO THE BARREL. IN SOME CASES, THIS PRINTING FOIL DUE TO A CLOG IN THE PRINTING STATION, HAS TO BE CUTE BY THE OPERATOR. WHEN THIS SITUATION HAPPENS, THE MACHINE STOPS AUTOMATICALLY, AND THE OPERATOR SHOULD CUT THE PRINTING FOIL REEL TO SOLVE THE CLOG. DURING THIS CUTTING PROCESS OF THE PRINTING FOIL, SOME PARTICLES COULD REMAIN IN THE MACHINE, AND IN THIS CASE, THESE FOIL PARTICLES FINALLY ENDED IN ONE SYRINGE BARREL, PRODUCING THE REPORTED ISSUE. THEREFORE, THE REPORTED NONCONFORMANCE IS CAUSED BY A DEFECTIVE FOIL REEL AND HUMAN ERROR. DHR SHOWED NO INDICATION OF THE ALLEGED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD SYRINGE S2 5ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: FM FOUND IN THE ORIGINAL PACKAGE. THE DEVICE HAS BEEN DESTROYED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD SYRINGE S2 5ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: FM FOUND IN THE ORIGINAL PACKAGE. THE DEVICE HAS BEEN DESTROYED.