BD SYRINGE S2 5ML
Report
- Report Number
- 3002682307-2019-00451
- Event Type
- Malfunction
- Date Received
- August 15, 2019
- Date of Event
- July 22, 2019
- Report Date
- August 21, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: BD HAS BEEN PROVIDED WITH A PHOTO FOR CATALOG 309050 LOT 1903206 TO INVESTIGATE FOR THIS RECORD. THE FOREIGN MATTER WAS IDENTIFIED AS PRINTING FOIL PARTICLES IN THE FLUID PATH. AS A RESULT, BD WAS ABLE TO VERIFY THE REPORTED ISSUE. THE PROCESS USED TO PRINT THE SCALE IN BD EMERALD SYRINGES IS CALLED "HOT STAMPING" PROCESS, IN WHICH, THROUGH A HEATED METAL STAMP, THE SCALE IS TRANSFERRED FROM THE PRINTING FOIL TO THE BARREL. IN SOME CASES, THIS PRINTING FOIL DUE TO A CLOG IN THE PRINTING STATION, HAS TO BE CUTE BY THE OPERATOR. WHEN THIS SITUATION HAPPENS, THE MACHINE STOPS AUTOMATICALLY, AND THE OPERATOR SHOULD CUT THE PRINTING FOIL REEL TO SOLVE THE CLOG. DURING THIS CUTTING PROCESS OF THE PRINTING FOIL, SOME PARTICLES COULD REMAIN IN THE MACHINE, AND IN THIS CASE, THESE FOIL PARTICLES FINALLY ENDED IN ONE SYRINGE BARREL, PRODUCING THE REPORTED ISSUE. THEREFORE, THE REPORTED NONCONFORMANCE IS CAUSED BY A DEFECTIVE FOIL REEL AND HUMAN ERROR. DHR SHOWED NO INDICATION OF THE ALLEGED DEFECT. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD SYRINGE S2 5ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: FM FOUND IN THE ORIGINAL PACKAGE. THE DEVICE HAS BEEN DESTROYED.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED WITH A BD SYRINGE S2 5ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: FM FOUND IN THE ORIGINAL PACKAGE. THE DEVICE HAS BEEN DESTROYED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 691128 | BD SYRINGE S2 5ML | SYRINGE | FMF | BECTON DICKINSON, S.A. | 1903206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |