SPINNING SPIROS® CLOSED MALE LUER, RED CAP
Report
- Report Number
- 9617594-2019-00269
- Event Type
- Malfunction
- Date Received
- August 14, 2019
- Date of Event
- July 30, 2019
- Report Date
- August 1, 2019
- Manufacturer
- ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
- Product Code
- FPA
- UDI-DI
- 00840619026615
- PMA / PMN Number
- K082806
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H10 - ONE USED LIST# CH2000S-C, SPINNING SPIROS® CLOSED MALE LUER, RED CAP. LOT # 4040735 AND ONE USED 35 ML SYRINGE, MANUFACTURER MONOJECT WERE RECEIVED FOR EVALUATION ON 8/9/2019. THE SPINNING SPIROS WAS PRESSURE LEAK TESTED AS PER THE PRODUCT PERFORMANCE SPECIFICATION AND NO LEAKAGE WAS OBSERVED AT ANY LOCATION WITHIN THE SPINNING SPIROS ASSEMBLY. THE SPIN FEATURE WAS SPINNING FREELY AND THE MATING DEVICE MALE AND FEMALE LUERS WERE ISO COMPATIBLE. THE COMPLAINT WAS UNABLE TO BE CONFIRMED. A DEVICE HISTORY REVIEW (DHR) FOR LOT# 4040735 AND RELEVANT COMMODITIES WERE REVIEWED. NO NON-CONFORMANCES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. D11 - UPDATED INFORMATION - 35ML SYRINGE, MFR MONOJECT ADDITIONAL INFORMATION CAN BE FOUND IN SECTION G1.
THE DEVICE IS EXPECTED TO RETURN FOR EVALUATION. IT HAS NOT BEEN RECEIVED.
THE EVENT INVOLVED A SPINNING SPIROS BEING THAT WHEN USED WITH AN UNSPECIFIED SYRINGE, A LEAK OF FLUOROURACIL WAS NOTED AROUND THE TIP OF THE SYRINGE AND THE CHEMO SPILLED ON THE PATIENT. THE SYRINGE WAS REMADE. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 688605 | SPINNING SPIROS® CLOSED MALE LUER, RED CAP | SET, ADMINISTRATION, INTRAVASCULAR | FPA | ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. | 4040735 | 00840619026615 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FLUOROURACIL, MFR UNK.| UNSP SYRINGE, UNK MFR. |