ADAPTPL 2 12HO TI
Report
- Report Number
- 8030965-2019-67219
- Event Type
- Malfunction
- Date Received
- August 14, 2019
- Report Date
- July 17, 2019
- Manufacturer
- OBERDORF SYNTHES PRODUKTIONS GMBH
- Product Code
- HRS
- UDI-DI
- 07611819071100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED: PART# 443.091 , LOT# 9875228 , MANUFACTURING SITE: MEZZOVICO, RELEASE TO WAREHOUSE DATE: 23. MARCH 2016. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. H3, H6: A PRODUCT INVESTIGATION WAS CONDUCTED. VISUAL INSPECTION: THE RECEIVED PLATE IS SLIGHTLY BENT AS REPORTED. THERE ARE WEAR MARKS AND SCRATCHES VISIBLE ON THE SURFACE OF THE PART. DOCUMENT/SPECIFICATION REVIEW: THE MANUFACTURING REVIEW SHOWS THAT THE PRODUCTION PROCEDURE WAS ACCORDING TO THE SPECIFICATIONS AND THERE WERE NO ISSUES THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. SUMMARY: THE RECEIVED CONDITION OF THE PLATE IS CONCORDANT WITH THE COMPLAINT DESCRIPTION AND THE COMPLAINT CONDITION IS CONFIRMED. THIS LOT WAS MANUFACTURED IN MARCH 2016 ACCORDING TO THE SPECIFICATION. BASED ON THAT AND THE CONDITION OF THE ITEM A PRODUCT RELATED ISSUE CAN BE EXCLUDED. SINCE THE PLATE SHOWS DAMAGES IN THE AREA WERE THE PLATE IS BENT, WE MUST ASSUME THAT THE PLATE WAS POSSIBLY TRIED TO BEND WITH A PLIERS. THE ROOT CAUSE WAS IDENTIFIED DURING THE PERFORMED CQ EVALUATION AND THEREFORE THE IN THE INVESTIGATION FLOW LISTED REMAINING INVESTIGATION STEPS ARE NOT REQUIRED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AS NO PRODUCT RELATED ISSUE COULD BE DETECTED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM HUNGARY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING DECONTAMINATION ON AN UNKNOWN DATE, THE PLATE WAS FOUND CURVED. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR AN ADAPTION PLATE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 689178 | ADAPTPL 2 12HO TI | PLATE, FIXATION, BONE | HRS | OBERDORF SYNTHES PRODUKTIONS GMBH | 9875228 | 07611819071100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |