FDA Adverse Event Malfunction Summary report: N

OXYGENATORS

MDR report key: 8891583 · Received August 14, 2019

Report

Report Number
8010762-2019-00244
Event Type
Malfunction
Date Received
August 14, 2019
Date of Event
July 13, 2019
Report Date
August 8, 2019
Manufacturer
MAQUET CARDIOPULMONARY GMBH
Product Code
DTZ
PMA / PMN Number
K082117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

MAQUET CARDIOPULMONARY AG IS AWARE OF SIMILAR COMPLAINTS FROM THIS PRODUCT. SAME PRODUCT, SHOWING A SIMILAR MALFUNCTION, HAS BEEN INVESTIGATED IN # (B)(4): WHEN PRIMING WITH WATER, A LEAK WAS DETECTED AT THE POINT CLAIMED BY THE CUSTOMER (POS 5.) 'OPENING FOR HOLDER'. SINCE FURTHER INVESTIGATION ON SAMPLE IS NOT POSSIBLE DUE TO SAFETY REASONS, THE CAUSE OF THE INCIDENT COULD NOT BE IDENTIFIED. IN ADDITION, MAQUET CARDIOPULMONARY GMBH IS AWARE OF SIMILAR FAILURE DESCRIPTION FOR SAME PRODUCT, COMPLAINT # (B)(4). THE VIDEO OF THIS FAILURE ALSO SHOWS THE OXYGENATOR IS LEAKING FROM HOUSING. THUS THE REPORTED FAILURE 'LEAK FROM THE BASE' COULD BE CONFIRMED. THIS COMPLAINT IS BEING MONITORED AS PART OF THE COMPLAINT DATA TRENDING OF MAQUET CARDIOPULMONARY GMBH AND FURTHER INVESTIGATIONS AND MEASURES WILL BE CONDUCTED IN CASE OF ADVERSE TRENDING.

Description of Event or Problem · 0

COMPLAINT#: (B)(4).

Additional Manufacturer Narrative · 1

A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

A LEAK WAS OBSERVED IN THE ADAPTER OF THE SUPPORT, BEING NECESSARY THE EXCHANGE OF ALL THE OXYGENATOR. ALL TESTS ON THE DEVICE WERE PERFORMED DURING ASSEMBLY AND THE CIRCULATORY PROCESS BEFORE CPB, AND WERE NOT ALTERED. AS THE PROBLEM OCCURRED MOMENTS BEFORE THE ENTRY INTO CPB, THERE WAS NO DAMAGE TO THE PATIENT, ONLY WAITING A FEW MINUTES FOR OXYGENATOR REPLACEMENT. (B)(4)..

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
687460 OXYGENATORS OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MAQUET CARDIOPULMONARY GMBH QUADROX-I ADULT (VKMO) W/ FILTER 92254623

Patients

Seq Age Sex Outcome Treatment
1