INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Report
- Report Number
- 3008452825-2019-00405
- Event Type
- Injury
- Date Received
- August 12, 2019
- Date of Event
- July 22, 2019
- Report Date
- August 12, 2019
- Manufacturer
- ST. JUDE MEDICAL, COSTA RICA LTDA
- Product Code
- DRF
- UDI-DI
- 05414734304146
- PMA / PMN Number
- K961924
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED DISSECTION COULD NOT BE CONCLUSIVELY DETERMINED. PER THE IFU, VASCULAR PERFORATION IS AN INHERENT RISK OF ANY ELECTRODE PLACEMENT.
RELATED MANUFACTURING REFERENCE: 3008452825-2019-00404, 3008452825-2019-00407, 2182269-2019-00113. DURING AN ATRIAL FIBRILLATION AND ATRIAL FLUTTER ABLATION PROCEDURE A DISSECTION OCCURRED. DURING THE PROCEDURE, AS THE PHYSICIAN WAS GUIDING THE DEVICES THROUGH A NARROW LEFT ILIAC VEIN, THE PATIENT BECAME HYPOTENSIVE. A HEMOGLOBIN TEST WAS RUN AND IT WAS DETERMINED THE PATIENT NEEDED A BLOOD TRANSFUSION. CONTRAST WAS INJECTED TO DETERMINE THE LOCATION OF THE BLEEDING AND A DISSECTION WAS FOUND. VASCULAR INTERVENTION WAS NOT REQUIRED AS THE DISSECTION CLOSED ON ITS OWN. THE PROCEDURE WAS COMPLETED AND THE PATIENT WAS STABLE. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 676206 | INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER | CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING | DRF | ST. JUDE MEDICAL, COSTA RICA LTDA | IBI-81404 | 6940899 | 05414734304146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention | INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER| LIVEWIRE¿ ELECTROPHYSIOLOGY CATHETER| VIEWFLEX¿ XTRA ICE CATHETER |