FDA Adverse Event Malfunction Summary report: N

PRISMAFLEX M100

MDR report key: 8883510 · Received August 12, 2019

Report

Report Number
8010182-2019-00167
Event Type
Malfunction
Date Received
August 12, 2019
Date of Event
July 11, 2019
Report Date
September 18, 2019
Manufacturer
BAXTER HEALTHCARE - MEYZIEU
Product Code
KDI
PMA / PMN Number
K041005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, PHOTOGRAPHS OF THE SAMPLE WERE PROVIDED FOR EVALUATION. THE VISUAL INSPECTION OF THE PROVIDED PICTURES SHOWED THAT ONE EXTREMITY OF THE PBP PUMP SEGMENT WAS DISCONNECTED FROM THE SUPPORT PLATE. THE PUMP SEGMENT WAS NOT INSERTED ALL THE WAY IN THE SUPPORT PLATE. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A MANUFACTURING ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER APPROXIMATELY ONE HOUR OF TREATMENT WITH A PRISMAFLEX SET AND A PRISMAFLEX CONTROL MACHINE, THE MACHINE ALARMED AN ¿AIR IN BLOOD ¿ AND THEN "GENERAL SYSTEM FAILURE" ALARM. FURTHER OBSERVATION REVEALED THAT THE PBP LINE WAS BROKEN AT THE JUNCTION OF THE PBP LINE AND THE CARTRIDGE. TREATMENT WAS DISCONTINUED AND A NEW SET WAS USED TO COMPLETE TREATMENT WITHOUT A REPORT OF BLOOD RETURN TO THE PATIENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
679921 PRISMAFLEX M100 DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI BAXTER HEALTHCARE - MEYZIEU NA 19A0201

Patients

Seq Age Sex Outcome Treatment
1 PRISMAFLEX CONTROL UNIT| PRISMAFLEX CONTROL UNIT