FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 8883360 · Received August 12, 2019

Report

Report Number
2518422-2019-01754
Event Type
Malfunction
Date Received
August 12, 2019
Date of Event
July 19, 2019
Report Date
July 19, 2019
Manufacturer
RESPIRONICS, INC
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED TEST STEPS DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678725 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC 1054260

Patients

Seq Age Sex Outcome Treatment
1