FDA Adverse Event Injury Summary report: N

BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 28 SIZE S -3.5

MDR report key: 8883103 · Received August 12, 2019

Report

Report Number
3005180920-2019-00683
Event Type
Injury
Date Received
August 12, 2019
Date of Event
July 18, 2019
Report Date
August 12, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805103
PMA / PMN Number
K072857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2019: LOT 177464 : (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27 FEBRUARY 2018. EXPIRATION DATE: 2023-02-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANTS INVOLVED BALL HEADS: BIPOLAR HEAD 25060.2850 BIPOLAR HEAD Ø28X50 LOT. 176898 (K091967): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01 MARCH 2018. EXPIRATION DATE: 2023-02-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY PERFORMED AFTER 2 MONTHS FROM THE PRIMARY DUE TO PAIN CAUSED BY FRACTURED ACETABULUM. PRIOR TO THE PRIMARY SURGERY, THE PATIENT HAD SEVERE BONE LOSS IN HER FEMORAL HEAD WHICH IN TURN CAUSED EROSION OF HER ACETABULUM. THE BIPOLAR HEAD UNDER STRESS PUSHED THROUGH THE ACETABULAR WALL. THE SURGEON REVISED BIPOLAR HEAD AND THE COCR HEAD SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
679793 BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 28 SIZE S -3.5 HIP COCR HEAD JDI MEDACTA INTERNATIONAL SA 177464 07630030805103

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention