BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 28 SIZE S -3.5
Report
- Report Number
- 3005180920-2019-00683
- Event Type
- Injury
- Date Received
- August 12, 2019
- Date of Event
- July 18, 2019
- Report Date
- August 12, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030805103
- PMA / PMN Number
- K072857
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON (B)(6) 2019: LOT 177464 : (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27 FEBRUARY 2018. EXPIRATION DATE: 2023-02-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANTS INVOLVED BALL HEADS: BIPOLAR HEAD 25060.2850 BIPOLAR HEAD Ø28X50 LOT. 176898 (K091967): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01 MARCH 2018. EXPIRATION DATE: 2023-02-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
REVISION SURGERY PERFORMED AFTER 2 MONTHS FROM THE PRIMARY DUE TO PAIN CAUSED BY FRACTURED ACETABULUM. PRIOR TO THE PRIMARY SURGERY, THE PATIENT HAD SEVERE BONE LOSS IN HER FEMORAL HEAD WHICH IN TURN CAUSED EROSION OF HER ACETABULUM. THE BIPOLAR HEAD UNDER STRESS PUSHED THROUGH THE ACETABULAR WALL. THE SURGEON REVISED BIPOLAR HEAD AND THE COCR HEAD SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 679793 | BALL HEADS: COCR COCR BALL HEAD 12/14 Ø 28 SIZE S -3.5 | HIP COCR HEAD | JDI | MEDACTA INTERNATIONAL SA | 177464 | 07630030805103 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |