MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET
Report
- Report Number
- 1820334-2019-01978
- Event Type
- Malfunction
- Date Received
- August 10, 2019
- Report Date
- September 16, 2019
- Manufacturer
- COOK INC
- Product Code
- DYB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. INVESTIGATION - EVALUATION. REVIEWS OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD, DRAWING, MANUFACTURER¿S INSTRUCTIONS, SPECIFICATIONS, AND QUALITY CONTROL PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. THE COMPLAINT DEVICE WAS NOT RETURNED; THEREFORE, NO PHYSICAL EXAMINATIONS COULD BE PERFORMED. AN ANNOTATED PHOTOGRAPH WAS PROVIDED BY THE USER FACILITY, HOWEVER, WHICH CLARIFIED THE POINT OF SEPARATION AT THE INNER HUB. A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. THERE IS NO EVIDENCE TO SUGGEST THE PRODUCT WAS NOT MADE TO SPECIFICATIONS. THE LOT NUMBER OF THE DEVICE IS NOT KNOWN; ACCORDINGLY, A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED. FURTHERMORE, REVIEWS OF THE MANUFACTURER¿S INSTRUCTIONS, DRAWINGS, SPECIFICATIONS, AND QUALITY CONTROL PROCEDURES WERE CONDUCTED, AND NO GAPS WERE DISCOVERED. BASED ON THE INFORMATION PROVIDED AND NO PRODUCT RETURNED, INVESTIGATION HAS CONCLUDED THAT NO CAUSE COULD BE ESTABLISHED FOR THE DEVICE FAILURE. WE WILL CONTINUE OUR MONITORING OF SIMILAR COMPLAINTS AND HAVE NOTIFIED THE APPROPRIATE PERSONNEL OF THIS EVENT. PER THE QUALITY ENGINEERING RISK ASSESSMENT, NO FURTHER ACTION IS REQUIRED. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
PMA/510(K) NUMBER: K171275. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.
AS REPORTED, DURING AN UNKNOWN PROCEDURE USING A MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET, THE HUB OF THE INNER CATHETER SEPARATED BETWEEN THE "UPPER" HUB AND THE LOCKING PORTION OF THE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 675649 | MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET | DYB INTRODUCER, CATHETER | DYB | COOK INC | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |