Narratives
Additional Manufacturer Narrative · 1
AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT GETTING INACCURATE READINGS ON HER PLINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READINGS (173) AS REF. POINT VS. BD READINGS (27, 232, 261) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.