FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 8878868 · Received August 9, 2019

Report

Report Number
3005985723-2019-00581
Event Type
Malfunction
Date Received
August 9, 2019
Date of Event
August 7, 2019
Report Date
February 27, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K142530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SMALL SCREW SHEARED OFF FROM INSIDE MICS HANDLE IN BETWEEN CUTTING ATTACHMENTS DURING TKA. ANGLED ATTACHMENT WOULD NOT SLIDE IN, THEN SMALL SHEARED SCREW FELL OUT ONTO FLOOR. CASE TYPE: TKA. SURGICAL DELAY: = 15 MINUTES. UPDATE: "NO DEBRIS LEFT BEHIND AFTER CASE AFTER STRAIGHT SAW CUTTING ATTACHMENT WAS REMOVED, COULD NOT GET THE ANGLED ATTACHMENT. EVENTUALLY HE WAS ABLE TO GET IT IN AND COMPLETED CASE. PIECE OF DEBRIS WAS FOUND ON FLOOR BEFORE END OF CASE SAW ATTACHMENT CLICKED AND LOCKED IN." PRODUCT INSPECTION: AS PER SERVICE MAXWO-01993212 & CASE NUMBER (B)(4). RMA# (B)(4). SN# (B)(6). FACTORY SERVICE TECHNICIAN: (B)(4). 1. UNABLE TO REPAIR MICS AS WITH THE NEW CABLE DID NOT FIX THE ISSUE. DISPOSITION - (RETURN TO VENDOR). OTHER (RTV) FAILED COLLAR TEST. THE ALLEGED FAILURE WAS CONFIRMED. DEVICE HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K09GT AND 24 DEVICES WERE ACCEPTED INTO FINAL STOCK ON 5/8/17. A REVIEW OF QT17-05-0034 REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 209063, LOT NUMBER 42030417 SHOWS 4 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSION: THE FAILURE MODE WAS CONFIRMED VIA INSPECTION. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED THEN THE COMPLAINT WILL BE REOPENED.

Description of Event or Problem · 0

SMALL SCREW SHEARED OFF FROM INSIDE MICS HANDLE IN BETWEEN CUTTING ATTACHMENTS DURING TKA. ANGLED ATTACHMENT WOULD NOT SLIDE IN, THEN SMALL SHEARED SCREW FELL OUT ONTO FLOOR. CASE TYPE: TKA. SURGICAL DELAY: = 15 MINUTES. UPDATE: "NO DEBRIS LEFT BEHIND AFTER CASE AFTER STRAIGHT SAW CUTTING ATTACHMENT WAS REMOVED, COULD NOT GET THE ANGLED ATTACHMENT. EVENTUALLY HE WAS ABLE TO GET IT IN AND COMPLETED CASE. PIECE OF DEBRIS WAS FOUND ON FLOOR BEFORE END OF CASE SAW ATTACHMENT CLICKED AND LOCKED IN."

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

SMALL SCREW SHEARED OFF FROM INSIDE MICS HANDLE IN BETWEEN CUTTING ATTACHMENTS DURING TKA. ANGLED ATTACHMENT WOULD NOT SLIDE IN, THEN SMALL SHEARED SCREW FELL OUT ONTO FLOOR CASE TYPE: TKA. SURGICAL DELAY: =15 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
672221 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 209063 4202103 / 42030417 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization