FDA Adverse Event Malfunction Summary report: N

PEN II OMNITROPE PEN 10

MDR report key: 8872075 · Received August 7, 2019

Report

Report Number
2243072-2019-01635
Event Type
Malfunction
Date Received
August 7, 2019
Date of Event
July 22, 2019
Report Date
August 21, 2019
Manufacturer
BECTON DICKINSON
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: ONE (1) SAMPLE WAS PROVIDED TO BD MEDICAL ¿ PHARMACEUTICAL SYSTEM (BDM-PS) FOR ANALYSIS. BDM-PS PERFORMED A BATCH HISTORY RECORD¿S REVIEW (BHR) INCLUDING A REVIEW OF ALL DATA COLLECTED DURING IN PROCESS AND QUALITY INSPECTIONS. THE BATCH INVOLVED IN THIS COMPLAINT MEETS ALL ACCEPTABLE QUALITY LEVELS (AQL¿S), WAS MANUFACTURED AND RELEASED ACCORDING TO APPLICABLE PROCEDURES AND SPECIFICATIONS. INITIAL EVALUATION OF THE COMPLAINT SAMPLE REVEALED A BROKEN VIAL RETAINER. THE ROOT CAUSE OF THE BROKEN VIAL RETAINER IS MOST LIKELY MATERIAL INCOMPATIBILITY. BREAKING FROM MATERIAL INCOMPATIBILITY IS DUE TO THE INTERACTION BETWEEN THE POLYCARBONATE OF THE VIAL RETAINER COMPONENT AND DIOCTYL PHTHALATE FOUND IN THE PEN POUCH. CHEMICAL COMPATIBILITY OVERVIEW FOR LEXAN POLYCARBONATE RECOMMENDS AGAINST THE USE OF DIOCYTL PHTHALATE IN CONJUNCTION WITH POLYCARBONATE AS IT RESULTS IN FAILURE OR SEVERE DEGRADATION. CORRECTIVE ACTIONS HAVE BEEN DETERMINED TO INFORM SANDOZ GMBH OF MATERIAL COMPATIBILITY WITH THE PEN POUCH. PREVENTATIVE ACTION HAS BEEN ESTABLISHED FOR SANDOZ TO UPDATE THE PEN POUCH MATERIAL. BASED ON ANALYSIS AND STUDIES THAT HAVE BEEN SUMMARIZED IN ¿SANDOZ OMNITROPE® PEN 5 & OMNITROPE® PEN 10 PEN CRACK ROOT CAUSE INVESTIGATION¿ (DATED 15-JUNE-2016) ONE POTENTIAL ROOT CAUSE HAS BEEN IDENTIFIED. 1.BREAKING ON THE VIAL RETAINER FROM MATERIAL INCOMPATIBILITY IS DUE TO THE INTERACTION BETWEEN THE POLYCARBONATE OF THE PEN COMPONENTS AND DIOCTYL PHTHALATE FOUND IN THE PEN POUCH. SANDOZ HAS BEEN INFORMED OF THE MATERIAL COMPATIBILITY ISSUE BY BD. SANDOZ HAS INFORMED BD THAT THE POUCH WILL BE UPDATED BY MIDDLE OF 2019. H3 OTHER TEXT : SEE SECTION H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PEN II OMNITROPE PEN 10 EXPERIENCED A FAILURE TO CONTAIN BLOOD/MEDICATION PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS STATED THAT: CONSUMER CALLED TO REPORT THAT HER PEN IS CRACKED. SHE SAYS THAT IT IS ONLY 2 MONTHS OLD. SHE DOES NOT HAVE THE PEN WITH HER.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. OEM MANUFACTURE: THE MANUFACTURING LOCATION FOR THIS PRODUCT IS SANDOZ. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AND THE (B)(4) FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PEN II OMNITROPE PEN 10 EXPERIENCED A FAILURE TO CONTAIN BLOOD/MEDICATION PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS STATED THAT: CONSUMER CALLED TO REPORT THAT HER PEN IS CRACKED. SHE SAYS THAT IT IS ONLY 2 MONTHS OLD. SHE DOES NOT HAVE THE PEN WITH HER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666480 PEN II OMNITROPE PEN 10 PEN FMI BECTON DICKINSON 17090003

Patients

Seq Age Sex Outcome Treatment
1 Other