REAGENT RED BLOOD CELLS BIOTESTCELL 1&2
Report
- Report Number
- 9610824-2019-00039
- Event Type
- Malfunction
- Date Received
- August 7, 2019
- Date of Event
- July 5, 2019
- Report Date
- August 7, 2019
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMBH
- Product Code
- KSZ
- UDI-DI
- 07611969952380
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS IS OUR COMBINED INITIAL AND FINAL REPORT ON THIS INCIDENT.
THE CUSTOMER REPORTED A FALSE NEGATIVE REACTION OF A PATIENT SAMPLE WHEN TESTED WITH BIOTESTCELL 1&2 ON TANGO INFINITY. THE SPECIFICITY OF THE MISSED ANTIBODY WAS ANTI-E. THE CUSTOMER STATED THAT THE ANTIBODY COULD BE DETECTED IN THE ORTHO GEL METHOD. THE CUSTOMER RETURNED THE PATIENT SAMPLE THAT HAD CAUSED A FALSE NEGATIVE TEST RESULT FOR INVESTIGATIONAL TESTING AND ALSO THE SUPPOSEDLY DEFECTIVE PRODUCT BIOTESTCELL 1&2. OUR QUALITY CONTROL LABORATORY TESTED THE PATIENT SAMPLE WITH THE COMPLAINT PRODUCT SAMPLE RETURNED BY THE CUSTOMER AND COULD CONFIRM THE CUSTOMER´S FINDING. FURTHER INVESTIGATIONS OF THE PATIENT SAMPLE WERE NOT POSSIBLE, BECAUSE IT WAS USED UP. THE RETURNED PRODUCT SAMPLE AND OUR QC LAB'S RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT WERE TESTED WITH DIFFERENT SAMPLES AND CONTROLS INCLUDING AN ANTI-E ON TANGO INFINITY. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE NEGATIVE REACTION. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THE CORRECT FUNCTION OF THE ALLEGEDLY DEFECTIVE LOT OF BIOTESTCELL 1&2. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVE INFLUENCES ON THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. REGARDING THE AFFECTED TANGO INFINITY: THE CUSTOMER PROVIDED RESULT IMAGES THAT SHOW THE NEGATIVE ANTIBODY SCREENING RESULT FOR AN ANTI-E, WHICH CAN BE ALSO VISUALLY CONFIRMED. LAST SEMI-ANNUAL PREVENTIVE MAINTENANCE WAS PERFORMED AS PER CHECKLIST ON APRIL 02, 2019. A FIELD SERVICE ENGINEER INSPECTED THE AFFECTED INSTRUMENT. ACCORDING TO THE FIELD SERVICE REPORT (WO REPORT) FROM JULY 11, 2019, ALL REQUESTED FILES WERE COLLECTED. THE PC WAS REBOOTED AND FIREWALL BOX TO RECONNECT BRICARE. IT WAS VERIFIED WITH A BRICARE ENGINEER THAT THE CONNECTION WAS WORKING. THE INSTRUMENT WAS OPERATING WITHIN MANUFACTURER SPECIFICATIONS AND RETURNED TO FULL SERVICE. THE LOG FILES DID NOT SHOW ANY ISSUE RELEVANT ABNORMALITIES. NO INDICATION FOR AN INSTRUMENT MALFUNCTION COULD BE IDENTIFIED ON CURRENT DATA. THE SERVICE ENGINEER CONFIRMED A PROPER FUNCTION OF THE INSTRUMENT. AS TO DATABASE, NO PROBLEMS WITH QUALITY CONTROL TESTING ON SOLIDSCREEN II METHOD ON THE DAYS OF ISSUE OCCURRED. THE CONTROL SAMPLES DID PASS EVERY DAY. THE INSTRUMENT GAVE REPEATEDLY THE SAME RESULTS AND THE REPORTED PROBLEM IS LIKELY RELATED TO SAMPLE SPECIFICITIES. THE ANTIBODIES WITHIN AFFECTED SAMPLE MAY HAVE BEEN PRESENT IN TOO LOW CONCENTRATION (BOARDERLINE) TO BE EVALUATED AS POSITIVE REACTION BY THE SOLIDSCREEN II METHOD. THE USAGE OF DIFFERENT METHODS MAY RESULT IN DIFFERENT RESULTS FOR THE ANTI-E ANTIBODY, BECAUSE IT IS A WEAK REACTING ANTIBODY AND THE METHODS HAVE DIFFERENT DETECTION SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 666604 | REAGENT RED BLOOD CELLS BIOTESTCELL 1&2 | BIOTESTCELL 1&2 | KSZ | BIO-RAD MEDICAL DIAGNOSTICS GMBH | 8923011-00 | 07611969952380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |