MAYFIELD 2000 SKULL CLAMP
Report
- Report Number
- 1525725-2004-00017
- Event Type
- Other
- Date Received
- October 29, 2004
- Report Date
- September 30, 2004
- Manufacturer
- OHIO MEDICAL INSTRUMENT CO
- Product Code
- HBF
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE SUSPECT DEVICE WAS RETURNED TO INTEGRA LIFESCIENCES CORP ON OCTOBER 1, 2004 FOR EVAL AND REPAIR. THE FOLLOWING IS A SUMMARY OF THE EVAL AND CONCLUSION OF RESULTS. EVAL: THE 80 POUND TORQUE KNOB TESTED IN THE ACCEPTABLE RANGE AT 20, 40 60, AND 80 POUNDS. THE LARGE STARBURST NEEDED HELI-COIL TREAD REPLACEMENT BECAUSE OF CROSS-THREADING. THE SWIVEL LOCK WAS DIFFICULT TO LOCK AND UNLOCK DUE TO A HEAVY BUILD-UP ON THE INTERNAL COMPONENTS. THE CAUTION LABEL WAS DAMAGED FROM CLEANING AT ELEVATED TEMPERATURES WHICH MAY HAVE ALSO DAMAGED THE INTERNAL COMPONENTS. AS REC'D, THE UNIT WAS TESTED UNDER PRESSURE AND WAS DIFFICULT TO PROPERLY POSITION, ADJUST AND LOCK. CONCLUSION: BASED ON THE NATURE OF THE INQUIRY, INTEGRA LIFESCIENCES EVAL RESULTS AND REPORTED INFO, NO DEFINITE CONCLUSION COULD BE DETERMINED AS TO HOW OR WHY THE DEVICE REPORTEDLY SLIPPED RESULTING IN A SCALP LACERATION. CORRECTIVE ACTION: NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. HOWEVER, AS AN AID TO THE NEURO STAFF AT MED CTR, INTEGRA LIFESCIENCES WILL SEND A COPY OF THE CD "MAYFIELD PROPER SKULL CLAMP POSITIONING". THE STAFF SHOULD FIND THIS HELPFUL.
WHILE POSITIONING HEADREST, SWIVEL ON THE DOUBLE PIN SEEMED TIGHT. WHEN PRESSURE WAS USED TO MOVE IT, THE PIN MOVED AND CREATED A LACERATION ON SCALP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAYFIELD 2000 SKULL CLAMP | SKULL CLAMP | HBF | OHIO MEDICAL INSTRUMENT CO | 4-O-A-2000 | 979 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Other |