Additional Manufacturer Narrative · 1

THE DEVICE IS NOT EXPECTED TO BE RETURNED. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFO.

Description of Event or Problem · 1

THE CUSTOMER HAS BEEN HAVING TROUBLE WITH INTEGRA'S 64 GRID (PART # 17194-8), MOST RECENTLY IN 2004. WHEN THE CUSTOMER CONNECTS THE PIG TAIL INTO THE CONNECTOR, THE SIGNAL IS WEAK OR NO SIGNAL WAS DETECTED AT ALL. WHEN A SIGNAL IS ACHIEVED, IT IS NOT VERY CLEAR COMPARED TO THE COMPETITOR AD-TECH PROD. ADD'L INFO WAS REC'D ON OCTOBER 6, 2004. THE PROBLEMS WERE N THE CONNECTIONS. THE LEADS DID NOT FIT INTO THE RECEPTACLE EASILY. THE RECORDINGS WERE UNINTERPRETABLE. ALL PROBLEMS WENT AWAY WHEN THE CUSTOMER REMOVED INTEGRA'S GRIDS AND REPLACED WITH COMPETITOR'S PROD. A RETURN IS NOT EXPECTED.