BLACKMAX-NEURO LOW PRESSURE
Report
- Report Number
- 1045834-2019-54246
- Event Type
- Malfunction
- Date Received
- August 6, 2019
- Report Date
- June 21, 2019
- Manufacturer
- DEPUY SYNTHES PRODUCTS LLC
- Product Code
- HBB
- UDI-DI
- 00845384001102
- PMA / PMN Number
- K831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
UDI: (B)(4). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING REPAIR, AN EVALUATION WAS PERFORMED AND IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER DEVICE USE WHICH IS USER ERROR. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MOTOR DEVICE HAD LOW POWER AND THE HOSE WAS DAMAGED / HAD A CUT. THE DEVICE ALSO FAILED PRETEST FOR INITIAL ROTATIONAL SPEED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS IN NEED OF REPAIR. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 659028 | BLACKMAX-NEURO LOW PRESSURE | MOTOR, DRILL, PNEUMATIC - HANDPIECE | HBB | DEPUY SYNTHES PRODUCTS LLC | 00845384001102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |