FDA Adverse Event Malfunction Summary report: N

BLACKMAX-NEURO LOW PRESSURE

MDR report key: 8864370 · Received August 6, 2019

Report

Report Number
1045834-2019-54246
Event Type
Malfunction
Date Received
August 6, 2019
Report Date
June 21, 2019
Manufacturer
DEPUY SYNTHES PRODUCTS LLC
Product Code
HBB
UDI-DI
00845384001102
PMA / PMN Number
K831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UDI: (B)(4). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING REPAIR, AN EVALUATION WAS PERFORMED AND IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER DEVICE USE WHICH IS USER ERROR. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MOTOR DEVICE HAD LOW POWER AND THE HOSE WAS DAMAGED / HAD A CUT. THE DEVICE ALSO FAILED PRETEST FOR INITIAL ROTATIONAL SPEED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS IN NEED OF REPAIR. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
659028 BLACKMAX-NEURO LOW PRESSURE MOTOR, DRILL, PNEUMATIC - HANDPIECE HBB DEPUY SYNTHES PRODUCTS LLC 00845384001102

Patients

Seq Age Sex Outcome Treatment
1