INRATIO
Report
- Report Number
- 2954730-2007-00342
- Event Type
- Malfunction
- Date Received
- July 27, 2007
- Date of Event
- July 17, 2007
- Report Date
- July 24, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE 2007, INRATIO - 5.3, LAB - 3.9, MEAN - 4.6, CONFIDENCE LIMITS - 2.6-6.9. INRATIO - 4.3, LAB - 3.9, MEAN - 4.1, CONFIDENCE LIMITS - 2.4-6.1. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070281. FIRST TEST INR = 5.3, SECOND TEST INR = 4.3, MEAN = 4.8; SD = 0.7; %CV = 15%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE 2007, INRATIO - 5.3, LAB - 3.9. INRATIO - 4.3, LAB - 3.9. CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: FIRST TEST INR = 5.3; SECOND TEST INR = 4.3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 070281 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |