FDA Adverse Event Malfunction Summary report: N

SYRINGE 1ML LL W/NDL ECLIPSE 27X1/2 RB

MDR report key: 8861640 · Received August 5, 2019

Report

Report Number
8041187-2019-00608
Event Type
Malfunction
Date Received
August 5, 2019
Date of Event
July 11, 2019
Report Date
August 30, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMI
UDI-DI
30382903057895
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: DHR WAS PERFORMED AND WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. ONE PHOTO AND ONE 1ML SYRINGE WITH AN ECLIPSE NEEDLE ASSEMBLY IN AN OPENED BLISTER PACK FROM BATCH 8052070 (P/N (B)(4)) WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THERE WAS MULTIPLE SCRATCHES AND INDENTATIONS ON AND BELOW THE FLANGE. THERE WAS ALSO A SCRATCH NEAR THE BOTTOM ABOVE THE COLLAR. IT WAS ALSO OBSERVED THE SCALE ON THE SYRINGE WAS SEVERELY SKEWED AND SMEARED TO THE POINT IT WAS ALMOST COMPLETELY ILLEGIBLE AND WAS REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE SKEWED SCALE DEFECT IS ASSOCIATED WITH THE PRINTING PROCESS. THIS WAS MOST LIKELY FROM THE PRINT PAD BEING IN THE WRONG POSITION CAUSING THE PRINT TO BE APPLIED AT AN ANGLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SCALE MARKING ISSUE OCCURRED BEFORE USE WITH A SYRINGE 1ML LL W/NDL ECLIPSE 27X1/2 RB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "SCALE MISSING".

Additional Manufacturer Narrative · 1

MEDICAL DEVICE TYPE: FMI; FMF. PMA / 510(K)#: K941562; K161170. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SCALE MARKING ISSUE OCCURRED BEFORE USE WITH A SYRINGE 1ML LL W/NDL ECLIPSE 27X1/2 RB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "SCALE MISSING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
656806 SYRINGE 1ML LL W/NDL ECLIPSE 27X1/2 RB HYPODERMIC SINGLE LUMEN NEEDLE;  PISTON SYRINGE FMI BECTON DICKINSON MEDICAL (SINGAPORE) 8052070 30382903057895

Patients

Seq Age Sex Outcome Treatment
1 Other