GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Report
- Report Number
- 2017233-2019-00598
- Event Type
- Injury
- Date Received
- August 5, 2019
- Date of Event
- March 1, 2019
- Report Date
- August 7, 2019
- Manufacturer
- W.L. GORE & ASSOCIATES
- Product Code
- NIO
- PMA / PMN Number
- P160021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL MANUFACTURER NARRATIVE: THE INITIAL MEDWATCH REPORT SUBMITTED UNDER MANUFACTURER REPORT NUMBER 2017233-2019-00598 WAS SUBMITTED IN ERROR, AND IS HEREBY RETRACTED. REFERENCE MEDWATCH #2017233-2019-00068.
REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE COMPLETED AS NO LOT NUMBER INFORMATION WAS PROVIDED. THE DEVICE WAS NOT RETURNED. CONSEQUENTLY, A DIRECT PRODUCT ANALYSIS WAS NOT POSSIBLE. ALL INFORMATION HAS BEEN PLACED ON FILE FOR USE IN TRACKING AND TRENDING.
THE FOLLOWING ARTICLE WAS REVIEWED: "FENESTRATED AORTIC ENDOGRAFT BRANCHING WITH GORE VBX POSES FAILURE RISK FROM DELAYED-ONSET BRANCH KINKING" (JENIANN A. YI, M.D.; DAVID KUWAYAMA, M.D.; JOURNAL OF VASCULAR SURGERY CASES AND INNOVATIVE TECHNIQUES, VOLUME 5, NUMBER 1; PUBLISHED BY ELSEVEIR INC.; HTTPS://DOI.ORG/10.1016/J.JVSCIT.2018.08.011). THE AIM OF THE STUDY WAS TO PRESENT A CASE OF NEAR OCCLUSION OF THE GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) CELIAC BRANCH DUE TO NARROWING OF THIS INTERSPACE IDENTIFIED POSTOPERATIVE MONTH 3. THE CELIAC ARTERY BRANCHING WAS PERFORMED USING THE VBX DEVICE. AT POST OP MONTH 3, MESENTERIC DUPLEX ULTRASOUND DEMONSTRATED ELEVATED VELOCITIES WITHIN THE VBX CONSISTENT WITH HIGH-GRADE STENOSIS. INTRAOPERATIVE IMAGING CONFIRMED SEVERE KINKING OF THE CELIAC STENT BRANCH. AFTER WIRE ACCESS WAS ESTABLISHED ACROSS THE STENOSIS, THE BRANCH WAS RELINED USING TWO ATRIUM ICAST STENTS. THE PATIENT TOLERATED THE PROCEDURE WELL AND WAS DISCHARGED HOME IN 24 HOURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655824 | GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS | STENT, ILIAC | NIO | W.L. GORE & ASSOCIATES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other |