LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Report
- Report Number
- 2937457-2019-02406
- Event Type
- Malfunction
- Date Received
- August 5, 2019
- Date of Event
- June 30, 2019
- Report Date
- August 5, 2019
- Manufacturer
- CONCORD MANUFACTURING
- Product Code
- FKX
- UDI-DI
- 00840861102068
- PMA / PMN Number
- K181108
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- 003
Narratives
PLANT INVESTIGATION: THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. AN EXTERIOR VISUAL INSPECTION OF THE RETURNED CYCLER SHOWED NO SIGNS OF PHYSICAL DAMAGE. UPON POWER UP, THE CYCLER TOUCH SCREEN TEST FAILED. WHEN POWERING ON THE CYCLER, THE OK, STOP AND UP/DOWN ARROW PUSH BUTTONS ILLUMINATED, HOWEVER THE FRONT PANEL TOUCH SCREEN REMAINED BLANK. AN INTERNAL INSPECTION OF THE CYCLER FOUND EVIDENCE OF AN INTERNAL SHORT PRESENT ON TRANSFORMER (T1) OF THE INVERTER BOARD. THE INVERTER BOARD IS LOCATED ON THE REAR OF THE FRONT PANEL ASSEMBLY. A KNOWN GOOD INVERTER BOARD WAS INSTALLED, AND THE TOUCH SCREEN BECAME OPERATIONAL. REMOVED FUNCTIONING INVERTER BOARD FROM THE TOUCH SCREEN AT THE COMPLETION OF THE INVESTIGATION. THERE WERE NO OTHER DISCREPANCIES ENCOUNTERED IN THE INTERNAL INSPECTION OF THE CYCLER. A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND VERIFIED THAT THE RESULTS OF THE IN-PROGRESS AND FINAL QUALITY CONTROL (QC) TESTING MET ALL REQUIREMENTS. UPON COMPLETION OF THE EVALUATION, THE REPORTED ISSUE WAS CONFIRMED, AND THE CAUSE WAS DETERMINED TO BE AN INTERNAL SHORT ON THE INVERTER BOARD. THE CYCLER WAS REFURBISHED FOLLOWING THE EVALUATION.
IT WAS REPORTED THAT THE SCREEN OF A PATIENT¿S LIBERTY SELECT CYCLER WENT BLANK DURING THEIR PERITONEAL DIALYSIS (PD) TREATMENT. THE CYCLER WAS PLUGGED INTO THE SOCKET PROPERLY. THE STATED THAT THE CYCLER CAN BE HEARD WORKING AND THE BUTTON IS LIGHT. AT THAT POINT IN TIME, THE TECHNICAL SUPPORT REPRESENTATIVE ADVISED THE PATIENT TO DISCONTINUE USE OF THE CYCLER AND TO NOTIFY THEIR PERITONEAL DIALYSIS REGISTERED NURSE (PDRN) OF THE EVENT. A REPLACEMENT CYCLER WAS ISSUED TO THE PATIENT. IT WAS REPORTED THAT AN ALTERNATE TREATMENT OPTION WAS AVAILABLE. UPON FOLLOW UP, THE PATIENT CONFIRMED THAT THERE WERE NO ADVERSE EVENTS OR MEDICAL INTERVENTION REQUIRED AS A RESULT OF THE REPORTED EVENT. THE PATIENT DID NOT COMPLETE TREATMENT. THE CYCLER WAS RETURNED TO THE MANUFACTURER AND A REPLACEMENT CYCLER WAS PROVIDED AND RECEIVED. UPON PHYSICAL EVALUATION OF THE CYCLER BY THE MANUFACTURER, IT WAS IDENTIFIED THAT THE TRANSFORMER ON THE INVERTER BOARD HAD AN INTERNAL SHORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654749 | LIBERTY SELECT CYCLER ASSY(NON-VALUATED) | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | CONCORD MANUFACTURING | 00840861102068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DELFLEX PD FLUID| LIBERTY CYCLER SET |