FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 8861342
·
Received August 5, 2019
Report
- Report Number
- 1314492-2019-02607
- Event Type
- Malfunction
- Date Received
- August 5, 2019
- Date of Event
- July 12, 2019
- Report Date
- August 5, 2019
- Manufacturer
- BAXTER HEALTHCARE - MEDINA
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BAXTER RECEIVED AND EVALUATED THE DEVICE. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THE REPORTED CONDITION WAS VERIFIED. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED DEVICE WOULD NOT RECOGNIZE BUBBLE WHICH WAS NOT REPRODUCED DURING EVALUATION. THE UPSTREAM/ULTRASONIC SENSOR WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP WOULD NOT RECOGNIZE BUBBLE. THE PROBLEM WAS FOUND IN THE BIOMED SERVICE DEPARTMENT DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654997 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE - MEDINA | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|