FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 8861342 · Received August 5, 2019

Report

Report Number
1314492-2019-02607
Event Type
Malfunction
Date Received
August 5, 2019
Date of Event
July 12, 2019
Report Date
August 5, 2019
Manufacturer
BAXTER HEALTHCARE - MEDINA
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAXTER RECEIVED AND EVALUATED THE DEVICE. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THE REPORTED CONDITION WAS VERIFIED. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED DEVICE WOULD NOT RECOGNIZE BUBBLE WHICH WAS NOT REPRODUCED DURING EVALUATION. THE UPSTREAM/ULTRASONIC SENSOR WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP WOULD NOT RECOGNIZE BUBBLE. THE PROBLEM WAS FOUND IN THE BIOMED SERVICE DEPARTMENT DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.