BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES
Report
- Report Number
- 1024879-2019-01351
- Event Type
- Malfunction
- Date Received
- August 5, 2019
- Date of Event
- July 18, 2019
- Report Date
- August 27, 2019
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- JKA
- UDI-DI
- 50382903679862
- PMA / PMN Number
- BK050036
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. ALTHOUGH NO SAMPLES OR PHOTOS WERE AVAILABLE FOR EVALUATION, BD HAS INITIATED FURTHER INVESTIGATION RELATING TO LABEL LIFT THROUGH A CAPA #1073079. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED.
IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES LABELS HAVE LIFTED OFF. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER COMPLAINT, BEFORE USE THIS BATCH, THEY FOUND MANY TUBES LABEL UPWARP.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES LABELS HAVE LIFTED OFF. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER COMPLAINT, BEFORE USE THIS BATCH, THEY FOUND MANY TUBES LABEL UPWARP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 656597 | BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO., (BD) | 9024657 | 50382903679862 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |