FDA Adverse Event Malfunction Summary report: N

BD MICRO FINE PLUS INSULIN PEN NEEDLE

MDR report key: 8859996 · Received August 5, 2019

Report

Report Number
9616656-2019-00704
Event Type
Malfunction
Date Received
August 5, 2019
Date of Event
July 19, 2019
Report Date
August 30, 2019
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: FOUR PHOTOS OF A 32G X 4MM PEN NEEDLE SAMPLE WERE RETURNED FROM LOT NO: 9022895, CAT NO: 320136. VISUAL EXAMINATION OF THE RETURNED PHOTOS WAS CARRIED OUT AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. BASED ON THE PHOTO RETURNED AND THE FACT NO SAMPLE WAS RETURNED FOR ANALYSIS IT IS NOT POSSIBLE TO CONFIRM THIS DEFECT TO BE MANUFACTURING RELATED. H3 OTHER TEXT.

Description of Event or Problem · 0

IT WAS REPORTED THAT MEDICATION WOULDN'T FLOW PROPERLY FROM THE BD MICRO FINE PLUS¿ INSULIN PEN NEEDLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "DRUG DIDN'T COME OUT PROPERLY."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MEDICATION WOULDN'T FLOW PROPERLY FROM THE BD MICRO FINE PLUS¿ INSULIN PEN NEEDLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "DRUG DIDN'T COME OUT PROPERLY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
656860 BD MICRO FINE PLUS INSULIN PEN NEEDLE PISTON SYRINGE FMF BECTON DICKINSON AND CO. 9022895

Patients

Seq Age Sex Outcome Treatment
1 Other