HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT
Report
- Report Number
- 1417411-2019-00035
- Event Type
- Malfunction
- Date Received
- August 2, 2019
- Date of Event
- November 12, 2018
- Report Date
- July 8, 2019
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- OGG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). ONE EMPTY 340 ML WATER BOTTLE LOT 220177 WAS RECEIVED EMPTY FOR EVALUATION WITH THE TRIGGER REMOVED AND TOP OF ADAPTOR PORT INTACT. IT CAME WITHOUT THE 040 HUMIDIFIER ADAPTOR. UPON VISUAL INSPECTION, THERE APPEARS TO BE A CHANNEL FROM THE BOTTOM OF THE ADAPTOR PORT TO THE TOP OF THE BOTTLE. THE BOTTLE WAS FILLED ABOUT 3/4 FULL WITH WATER AND AN ADAPTOR (LOT 17F18) WAS ATTACHED. THE WATER BOTTLE/ADAPTOR ASSEMBLY WAS ATTACHED TO A FLOWMETER. NO BUBBLES WERE OBSERVED AT 0.5 LITERS PER MINUTE OR AT 1 LITER PER MINUTE. UPON INCREASING THE AIR FLOW BUBBLES APPEARED IN THE BOTTLE. A REVIEW OF THE MANUFACTURING EVENT LOG OF THE LOT NUMBER REPORTED SHOWS NO ISSUES THAT MAY HAVE CONTRIBUTED TO ANY QUALITY ISSUES REPORTED. ALL PROCESS PARAMETERS WERE WITHIN SPECIFICATION AND ALL IN-PROCESS QA INSPECTIONS WERE ACCEPTABLE. BASED ON THE INVESTIGATION PERFORMED, THE REPORTED COMPLAINT IS CONFIRMED. CORRECTIVE ACTION HAS BEEN TAKEN. THE RETURNED SAMPLE WAS MANUFACTURED PRIOR TO THE CORRECTIVE ACTION.
CUSTOMER COMPLAINT STATES: "PRESENCE OF A BUBBLE IN THE PLASTIC PART THAT CONNECTS THE PLUG IN OF THE OXYGEN TO THE AQUAPACK BOTTLE. SO IT OCCLUDES THE FLOW OF OXYGEN." THE CONDITION OF THE PATIENT IS REPORTED AS "FINE".
(B)(4).
CUSTOMER COMPLAINT STATES: "PRESENCE OF A BUBBLE IN THE PLASTIC PART THAT CONNECTS THE PLUG IN OF THE OXYGEN TO THE AQUAPACK BOTTLE. SO IT OCCLUDES THE FLOW OF OXYGEN." THE CONDITION OF THE PATIENT IS REPORTED AS "FINE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 649948 | HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT | HUMIDIFIER NEBULIZER KIT | OGG | TELEFLEX MEDICAL | 220177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NONE REPORTED.| NONE REPORTED. |