FDA Adverse Event Malfunction Summary report: N

HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT

MDR report key: 8856838 · Received August 2, 2019

Report

Report Number
1417411-2019-00034
Event Type
Malfunction
Date Received
August 2, 2019
Date of Event
November 12, 2018
Report Date
July 8, 2019
Manufacturer
TELEFLEX MEDICAL
Product Code
OGG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

QN#(B)(4). ONE EMPTY 340 ML WATER BOTTLE LOT 220177 WAS RECEIVED EMPTY FOR EVALUATION WITH THE TRIGGER REMOVED AND TOP OF ADAPTOR PORT INTACT. IT CAME WITHOUT THE 040 HUMIDIFIER ADAPTOR. UPON VISUAL INSPECTION, THERE APPEARS TO BE A CHANNEL FROM THE BOTTOM OF THE ADAPTOR PORT TO THE TOP OF THE BOTTLE. THE BOTTLE WAS FILLED ABOUT 3/4 FULL WITH WATER AND AN ADAPTOR (LOT 17F18) WAS ATTACHED. THE WATER BOTTLE/ADAPTOR ASSEMBLY WAS ATTACHED TO A FLOWMETER. NO BUBBLES WERE OBSERVED AT 0.5 LITERS PER MINUTE OR AT 1 LITER PER MINUTE. UPON INCREASING THE AIR FLOW TO ABOUT 3 LITERS PER MINUTE, BUBBLES APPEARED IN THE BOTTLE. A REVIEW OF THE MANUFACTURING EVENT LOG OF THE LOT NUMBER REPORTED SHOWS NO ISSUES THAT MAY HAVE CONTRIBUTED TO ANY QUALITY ISSUES REPORTED. ALL PROCESS PARAMETERS WERE WITHIN SPECIFICATION AND ALL IN-PROCESS QA INSPECTIONS WERE ACCEPTABLE. BASED ON THE INVESTIGATION PERFORMED, THE REPORTED COMPLAINT IS CONFIRMED. CORRECTIVE ACTION HAS BEEN TAKEN. THE RETURNED SAMPLE WAS MANUFACTURED PRIOR TO THE CORRECTIVE ACTION.

Description of Event or Problem · 0

CUSTOMER COMPLAINT STATES: "PRESENCE OF A BUBBLE IN THE PLASTIC PART THAT CONNECTS THE PLUG IN OF THE OXYGEN TO THE AQUAPACK BOTTLE. SO IT OCCLUDES THE FLOW OF OXYGEN." THE CONDITION OF THE PATIENT IS REPORTED AS "FINE".

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

CUSTOMER COMPLAINT STATES: "PRESENCE OF A BUBBLE IN THE PLASTIC PART THAT CONNECTS THE PLUG IN OF THE OXYGEN TO THE AQUAPACK BOTTLE. SO IT OCCLUDES THE FLOW OF OXYGEN." THE CONDITION OF THE PATIENT IS REPORTED AS "FINE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650712 HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT HUMIDIFIER NEBULIZER KIT OGG TELEFLEX MEDICAL 220177

Patients

Seq Age Sex Outcome Treatment
1 NONE REPORTED.| NONE REPORTED.