FDA Adverse Event Malfunction Summary report: N

MEGADYNE ACE BLADE 700 ZIP E

MDR report key: 8855738 · Received August 2, 2019

Report

Report Number
1721194-2019-00106
Event Type
Malfunction
Date Received
August 2, 2019
Date of Event
July 16, 2019
Report Date
July 16, 2019
Manufacturer
MEGADYNE MEDICAL PRODUCTS, INC.
Product Code
GEI
PMA / PMN Number
K141587
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DATE SENT: 5/5/2020 INVESTIGATION SUMMARY WE FOUND THAT THERE WAS A MISSING CORNER AT THE TIP OF THE BLADE AND THERE WAS A TRACE OF BURNING. WE SIMULATED THE FIRING TEST WITH THE SAME MODEL PRODUCT (SETTING CONDITION: CUT 50 WATT & COAG 50 WATT), AFTER 100 TIMES, IT DID NOT PRODUCE THE SAME CONDITION AS THE CUSTOMER COMPLAINT SAMPLE. MOREOVER, WE DIDN'T RECEIVE FURTHER INFORMATION ABOUT THE DEVICE DEFECTIVE SITUATION FROM THE END-USER. OUR TEAM WILL NEED MORE INFORMATION FEEDBACK FROM THE CLINICAL SPECIALIST WITH REGARDS TO PRODUCT USAGE CONDITIONS (E.G. THE DETAILED BACKGROUND OF THE INCIDENT, MATING DEVICE, ETC.). THEREFORE, NEW DEANTRONICS COULD NOT IDENTIFY THE ROOT CAUSE OF THIS CUSTOMER COMPLAINT. THE BATCH/LOT NUMBER DOES NOT MATCH AND CANNOT PERFORM THE DHR REVIEW.

Additional Manufacturer Narrative · 1

(B)(4). DATE SENT: 8/2/2019. BATCH # UNK. FOLLOWING INFORMATION HAS BEEN REQUESTED BUT UNAVAILABLE: PLEASE CLARIFY: ¿CORNER PART OF THE BLADE CHIPPED OFF¿ DID THE CORNER OF THE ACTUAL BLADE BREAK OFF? DID THE COATING ON THE CORNER OF THE BLADE FLAKE OFF? ARE THERE ANY PLANS TO ATTEMPT TO LOCATE THE MISSING FRAGMENT?

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A BREAST MASTOPEXY A CORNER PART OF THE BLADE CHIPPED OFF, GENERATING A FRAGMENT. IT IS UNKNOWN IF THE FRAGMENT WAS FOUND. A SIMILAR DEVICE WAS USED. THE PROCEDURE WAS NOT COMPLETE AT TIME OF SUBMISSION. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651143 MEGADYNE ACE BLADE 700 ZIP E MEGADYNE ACE 700 ZIP PEN E CONNECTOR GEI MEGADYNE MEDICAL PRODUCTS, INC.

Patients

Seq Age Sex Outcome Treatment
1