FDA Adverse Event Death Summary report: N

GE OEC 9400

MDR report key: 885261 · Received July 26, 2007

Report

Report Number
1720753-2007-03535
Event Type
Death
Date Received
July 26, 2007
Date of Event
July 1, 2007
Report Date
July 26, 2007
Manufacturer
GE OEC MEDICAL SYSTEMS INC
Product Code
JAA
PMA / PMN Number
K864837
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS SYSTEM IS PAST THE END OF LIFE CYCLE AND IS OBSOLETE. NO PARTS FOR REPAIR ARE AVAILABLE. ALL CUSTOMERS WITH OEC SERIES 9400 SYSTEMS WERE NOTIFIED OF SYSTEM END OF LIFE EFFECTIVE DECEMBER 31, 2003. A GE OEC SERVICE REPRESENTATIVE INVESTIGATED THE ISSUE AND INFORMED THE CUSTOMER THAT THE BATTERIES WERE THE LIKELY CAUSE OF THE IMAGE QUALITY ISSUE NOTED FOR THE SYSTEM BEING REMOVED DURING THE PROCEDURE. BATTERIES FOR THIS SYSTEM ARE NO LONGER AVAILABLE AS THE SYSTEM IS OBSOLETE, AND SERVICE COULD NOT BE PERFORMED. THE SYSTEM, EVEN IF OBSOLETE, REQUIRES THAT IT BE PLUGGED IN, IF IT IS NOT USED, OR IT IS STORED FOR A PROLONGED PERIOD OF TIME (3 MONTHS OR LONGER). IT IS INDICATED, IN THE SERVICE REPORT, THAT THE SYSTEM MAY HAVE BEEN STORED FOR AN EXTENDED PERIOD OF TIME, WITHOUT BEING PLUGGED IN, AND THIS WOULD NEGATIVELY AFFECT THE BATTERIES AND SYSTEM PERFORMANCE, INCLUDING IMAGE QUALITY.

Description of Event or Problem · 1

THE GE OEC 9400 FLUOROSCOPY SYSTEM WAS REPORTED TO HAVE POOR IMAGE QUALITY. THE SYSTEM WAS BEING USED FOR A TRANSCUTANEOUS PACEMAKER INSERTION IN THE FACILITY INTENSIVE CARE UNIT. IT WAS REPORTED THAT THE PACEMAKER LEAD COULD NOT BE VISUALIZED BY THIS SYSTEM AND IT WAS REMOVED FROM THE BEDSIDE FOR REPLACEMENT WITH ANOTHER C-ARM TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THE PT EXPIRED PRIOR TO THE REPLACEMENT C-ARM ARRIVAL IN THE INTENSIVE CARE UNIT. REPORTED ISSUE FROM FACILITY: POOR IMAGE QUALITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9400 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC GE OEC 9400 NA

Patients

Seq Age Sex Outcome Treatment
1 76 YR Death