FDA Adverse Event Malfunction Summary report: N

OXYGENATORS

MDR report key: 8850353 · Received August 1, 2019

Report

Report Number
8010762-2019-00231
Event Type
Malfunction
Date Received
August 1, 2019
Date of Event
July 4, 2019
Report Date
September 13, 2019
Manufacturer
MAQUET CARDIOPULMONARY GMBH
Product Code
DTZ
PMA / PMN Number
K082117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

MAQUET CARDIOPULMONARY GMBH WAS NOT REQUESTED THE PRODUCT BACK FOR INVESTIGATION SINCE THE ISSUE IS KNOWN. MAQUET CARDIOPULMONARY AG IS AWARE OF SIMILAR COMPLAINTS FROM THIS PRODUCT. SIMILAR PRODUCT, SHOWING A SIMILAR MALFUNCTION, HAS BEEN INVESTIGATED IN (B)(4) . UNDER SCOPE OF THE COMPLAINT (B)(4) ; MAQUET CARDIOPULMONARY GMBH WAS RECEIVED THE PRODUCT BACK FOR INVESTIGATION. A QUADROX-I WAS RETURNED. THE PRODUCT WAS CONTAMINATED. VISUAL INSPECTION WAS PERFORMED. A LONG CRACK ON THE LUER LOCK NEAR THE VENT COCK WAS DETECTED, SO, IT WAS DECIDED THAT THE LEAK TEST WAS NOT NECESSARY. NO OTHER ABNORMALITIES COULD BE DETECTED. THE FAILURE WAS CONFIRMED BY THE LABORATORY. THE MANUFACTURER`S REVIEW OF THE CONTROL FORM IN PRODUCTION INDICATED THAT A 100% VISUAL INSPECTION IS CONDUCTED DURING MANUFACTURING. THE INFORMATION OBTAINED SO FAR IN THIS INVESTIGATION WOULD CONFIRM THAT THE DEVICE MET ITS SPECIFICATION AT THE TIME OF MANUFACTURING AND THEREFORE ALL DAMAGES FOUND ON THE PRODUCT ARE DUE TO EXCESSIVE OR INADEQUATE EXTERNAL PHYSICAL FORCE THAT WAS EXERTED ON THE PRODUCT AFTER THE RELEASE.

Description of Event or Problem · 0

COMPLAINT: #(B)(4).

Additional Manufacturer Narrative · 1

A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. IMPORTER- (B)(4).

Description of Event or Problem · 1

BLEED COCK AT THE INLET OF THE INTEGRATED FILTER DRIPS. COMPLAINT NO: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
646805 OXYGENATORS OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MAQUET CARDIOPULMONARY GMBH QUADROX-I ADULT (VKMO) W/ FILTER

Patients

Seq Age Sex Outcome Treatment
1