FDA Adverse Event Malfunction Summary report: N

ACCUTREND CHOLESTEROL

MDR report key: 8846998 · Received July 31, 2019

Report

Report Number
1823260-2019-02791
Event Type
Malfunction
Date Received
July 31, 2019
Date of Event
July 9, 2019
Report Date
August 22, 2019
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CHH
PMA / PMN Number
K051376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER'S METER AND STRIPS WERE RETURNED FOR INVESTIGATION. THE RETURNED PRODUCT WAS MEASURED IN COMPARISON TO A REFERENCE METER AND RETENTION LOT STRIPS. CUSTOMER SAMPLES CHOLESTEROL LOT 349024-04 AND RETENTION SAMPLES CHOLESTEROL LOT 349024-04 WERE MEASURED WITH CONTROLS LEVEL LOW /HIGH LOT 403614-01 ON CUSTOMER ACCUTREND PLUS DEVICES AND REFERENCE DEVICES. LOW CONTROL MEASUREMENTS: TESTING RESULTS (QC RANGE: 163 - 207 MG/DL): CUSTOMER METER: QC 1: 195.0 MG/DL, QC 2: 188.1 MG/DL, QC 3: 193.9 MG/DL. REFERENCE METER: QC 1: 193.0 MG/DL, QC 2: 181.0 MG/DL, QC 3: 188.0 MG/DL HIGH CONTROL MEASUREMENTS: TESTING RESULTS (QC RANGE: 224 - 284 MG/DL): CUSTOMER METER: QC 1: 246.1 MG/DL, QC 2: 257.0 MG/DL, QC 3: 248.8 MG/DL. REFERENCE METER: QC 1: 251.0 MG/DL, QC 2: 260.0 MG/DL, QC 3: 268.0 MG/DL. ALL RESULTS WERE WITHIN THE CONTROL RANGE. THE RESULTS OF ALL MEASUREMENTS FULFILLED THE REQUIREMENTS. NO SIGNIFICANT DIFFERENCES COULD BE IDENTIFIED BETWEEN THE CUSTOMER SAMPLES /CUSTOMER METER AND REFERENCE DEVICES/ RETENTION SAMPLES.

Additional Manufacturer Narrative · 1

QUALITY CONTROL RESULTS WERE ACCEPTABLE. THE CUSTOMER¿S STRIPS WERE REQUESTED FOR RETURN. THE PRODUCT HAS NOT BEEN RECEIVED AT THIS TIME. IF THE PRODUCT IS RETURNED IN THE FUTURE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE EVENT OCCURRED IN: (B)(6).

Description of Event or Problem · 1

THE INITIAL REPORTER COMPLAINED OF A QUESTIONABLE CHOLESTEROL RESULT FROM AN ACCUTREND PLUS METER SERIAL NUMBER (B)(4) COMPARED TO AN UNSPECIFIED LABORATORY METHOD. THE INITIAL CHOLESTEROL RESULT FROM THE METER WAS LOW. A "LOW" ALARM MEANS THE RESULT WAS BELOW THE MEASUREMENT RANGE. THE MEASUREMENT RANGE FOR CHOLESTEROL IS 3.88 - 7.76 MMOL/L. THE LABORATORY RESULT WAS 7 MMOL/L. THE RESULT IN QUESTION WAS REPORTED OUTSIDE OF THE LABORATORY.