ACCUTREND CHOLESTEROL
Report
- Report Number
- 1823260-2019-02791
- Event Type
- Malfunction
- Date Received
- July 31, 2019
- Date of Event
- July 9, 2019
- Report Date
- August 22, 2019
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- CHH
- PMA / PMN Number
- K051376
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER'S METER AND STRIPS WERE RETURNED FOR INVESTIGATION. THE RETURNED PRODUCT WAS MEASURED IN COMPARISON TO A REFERENCE METER AND RETENTION LOT STRIPS. CUSTOMER SAMPLES CHOLESTEROL LOT 349024-04 AND RETENTION SAMPLES CHOLESTEROL LOT 349024-04 WERE MEASURED WITH CONTROLS LEVEL LOW /HIGH LOT 403614-01 ON CUSTOMER ACCUTREND PLUS DEVICES AND REFERENCE DEVICES. LOW CONTROL MEASUREMENTS: TESTING RESULTS (QC RANGE: 163 - 207 MG/DL): CUSTOMER METER: QC 1: 195.0 MG/DL, QC 2: 188.1 MG/DL, QC 3: 193.9 MG/DL. REFERENCE METER: QC 1: 193.0 MG/DL, QC 2: 181.0 MG/DL, QC 3: 188.0 MG/DL HIGH CONTROL MEASUREMENTS: TESTING RESULTS (QC RANGE: 224 - 284 MG/DL): CUSTOMER METER: QC 1: 246.1 MG/DL, QC 2: 257.0 MG/DL, QC 3: 248.8 MG/DL. REFERENCE METER: QC 1: 251.0 MG/DL, QC 2: 260.0 MG/DL, QC 3: 268.0 MG/DL. ALL RESULTS WERE WITHIN THE CONTROL RANGE. THE RESULTS OF ALL MEASUREMENTS FULFILLED THE REQUIREMENTS. NO SIGNIFICANT DIFFERENCES COULD BE IDENTIFIED BETWEEN THE CUSTOMER SAMPLES /CUSTOMER METER AND REFERENCE DEVICES/ RETENTION SAMPLES.
QUALITY CONTROL RESULTS WERE ACCEPTABLE. THE CUSTOMER¿S STRIPS WERE REQUESTED FOR RETURN. THE PRODUCT HAS NOT BEEN RECEIVED AT THIS TIME. IF THE PRODUCT IS RETURNED IN THE FUTURE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE EVENT OCCURRED IN: (B)(6).
THE INITIAL REPORTER COMPLAINED OF A QUESTIONABLE CHOLESTEROL RESULT FROM AN ACCUTREND PLUS METER SERIAL NUMBER (B)(4) COMPARED TO AN UNSPECIFIED LABORATORY METHOD. THE INITIAL CHOLESTEROL RESULT FROM THE METER WAS LOW. A "LOW" ALARM MEANS THE RESULT WAS BELOW THE MEASUREMENT RANGE. THE MEASUREMENT RANGE FOR CHOLESTEROL IS 3.88 - 7.76 MMOL/L. THE LABORATORY RESULT WAS 7 MMOL/L. THE RESULT IN QUESTION WAS REPORTED OUTSIDE OF THE LABORATORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 640695 | ACCUTREND CHOLESTEROL | CHOLESTEROL (TOTAL) TEST SYSTEM | CHH | ROCHE DIAGNOSTICS | NA | 34902404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |