FDA Adverse Event Malfunction Summary report: N

BD NEEDLE CLIPPING DEVICE SAFE CLIP

MDR report key: 8846645 · Received July 31, 2019

Report

Report Number
2243072-2019-01541
Event Type
Malfunction
Date Received
July 31, 2019
Date of Event
July 15, 2019
Report Date
September 17, 2019
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
00382903282357
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. ACCORDING WITH THE DHR REVIEW INFORMATION, THE PROBLEM ¿NO CLIPPING¿ HAS NO NHB ASSEMBLY PROCESS RELATION, ALL SAMPLES OF SAFE CLIP DISPOSABLE CUTTER (USA) PASSED FUNCTIONAL TEST. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SAFECLIP IS NOT CLIPPING THE NEEDLE WITH A BD NEEDLE CLIPPING DEVICE SAFE CLIP. THIS OCCURRED ON 3 SEPARATE OCCASIONS, HOWEVER, THE DATE/TIME IS UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT CONSUMER'S SAFECLIP IS NOT CLIPPING ANY NEEDLE.

Additional Manufacturer Narrative · 1

FOR OEM MANUFACTURING SITES: IN THIS MDR, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED AS NYPRO, MX IS AN OEM MANUFACTURING SITE. MEDICAL DEVICE EXPIRATION DATE: N/A. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SAFE CLIP IS NOT CLIPPING THE NEEDLE WITH A BD NEEDLE CLIPPING DEVICE SAFE CLIP. THIS OCCURRED ON 3 SEPARATE OCCASIONS, HOWEVER, THE DATE/TIME IS UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT CONSUMER'S SAFE CLIP IS NOT CLIPPING ANY NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640817 BD NEEDLE CLIPPING DEVICE SAFE CLIP PISTON SYRINGE FMF BECTON DICKINSON 8288025 00382903282357

Patients

Seq Age Sex Outcome Treatment
1 Other