FDA Adverse Event Malfunction Summary report: N

BD SYRINGE SOLOMED 2PC 2ML 25X5/8 FIXED CN

MDR report key: 8846642 · Received July 31, 2019

Report

Report Number
3002682307-2019-00423
Event Type
Malfunction
Date Received
July 31, 2019
Date of Event
June 27, 2019
Report Date
August 29, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAS NOT BEEN PROVIDED WITH PHOTOS OR SAMPLES FOR CATALOG 302772 LOT 1801406 TO INVESTIGATE FOR THIS RECORD. BD WAS UNABLE TO VERIFY THE REPORTED ISSUE OR DETERMINE A DEFINITIVE ROOT CAUSE. THE 20 REFERENCE UNITS THAT HAVE BEEN RETAINED AND ANALYZED FOR THIS RECORD SHOWED NO DIFFICULTIES OR DEFECTS. IN ADDITION, THE PACKAGE INTEGRITY HAS BEEN WITHIN SPECIFICATIONS. AS THIS IS THE FIRST TIME THIS LOT HAS BEEN REPORTED, NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. DHR: BD HAS REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QN NOR NCMR'S H3 OTHER TEXT .

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE IT WAS DISCOVERED THAT STERILITY WAS COMPROMISED DUE TO POOR SEAL INTEGRITY WITH A BD SYRINGE SOLOMED 2PC 2ML 25X5/8 FIXED CN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: BEFORE USE, IT'S NOTICED THAT THE PACKAGE SEAL INTEGRITY POOR.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE #: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE IT WAS DISCOVERED THAT STERILITY WAS COMPROMISED DUE TO POOR SEAL INTEGRITY WITH A BD SYRINGE SOLOMED 2PC 2ML 25X5/8 FIXED (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: BEFORE USE, IT'S NOTICED THAT THE PACKAGE SEAL INTEGRITY POOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640819 BD SYRINGE SOLOMED 2PC 2ML 25X5/8 FIXED CN SYRINGE FMF BECTON DICKINSON, S.A. 1801406

Patients

Seq Age Sex Outcome Treatment
1 Other