FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 8846165 · Received July 31, 2019

Report

Report Number
2031642-2019-05277
Event Type
Malfunction
Date Received
July 31, 2019
Report Date
July 18, 2019
Manufacturer
RESPIRONICS CALIFORNIA, INC
Product Code
MNT
UDI-DI
00884838020054
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: (B)(6) 2019. DATE OF REPORT: 31JULY2019. THE AC POWER CORD (POWER CORD,3 WIRE,RT ANGLE,NA,10A)WAS RETURNED TO THE FI (FAILURE INVESTIGATION) LAB FOR EVALUATION. NO OBVIOUS WEAR, DAMAGE OR CUTS ALONG MAIN CONDUCTOR INSULATION. INSPECTION OF MAIN CORD INSULATOR TO BOTH PLUG END STRAIN RELIEFS REVEALED SEPARATION OF INSULATOR FROM WALL PLUG INSIDE DIAMETER. THE CUSTOMER REPLACED THE DEFECTIVE AC POWER CORD TO ADDRESS THE REPORTED PROBLEM. THE UNIT SUCCESSFULLY PASSED THE REQUIRED PERFORMANCE VERIFICATION TEST. VERIFIED REPORTED FAILURE OF ESA RESISTANCE OUT OF SPECIFICATION. VERIFIED GND-GND RESISTANCE OUT OF SPECIFICATION. NOTED RESISTANCE OF 775KOHMS BETWEEN GND-L. ROOT CAUSE DETERMINED TO BE MECHANICAL DAMAGE OF CABLE AT WALL PLUG END.

Description of Event or Problem · 1

THE CUSTOMER REPORTED UNIT FAILED ELECTRICAL TEST, POWER CORD RESISTANCE AND FAILED THE (PREVENTATIVE MAINTENANCE) PM. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638682 V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE MNT RESPIRONICS CALIFORNIA, INC V60 00884838020054

Patients

Seq Age Sex Outcome Treatment
1