FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 8846062 · Received July 31, 2019

Report

Report Number
1024879-2019-01308
Event Type
Malfunction
Date Received
July 31, 2019
Date of Event
July 15, 2019
Report Date
September 17, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS AND SAMPLE WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR EMBEDDED FOREIGN MATERIAL IN THE TUBE WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE EXPERIENCED FOREIGN MATTER CONTAMINATION NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 367986, BATCH NO: 9081742. WE HAVE IDENTIFIED A LOT# OF GOLD TOP TUBES, WHICH HAVE BLACK SPECKS INSIDE THE TUBE.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE EXPERIENCED FOREIGN MATTER CONTAMINATION NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 367986. BATCH NO: 9081742. WE HAVE IDENTIFIED A LOT# OF GOLD TOP TUBES, WHICH HAVE BLACK SPECKS INSIDE THE TUBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639583 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 9081742 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other