FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1712K

MDR report key: 8844520 · Received July 31, 2019

Report

Report Number
2032227-2019-38582
Event Type
Malfunction
Date Received
July 31, 2019
Date of Event
July 22, 2019
Report Date
March 16, 2020
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169999770
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

DEVICE WAS RECEIVED WITH CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM DURING TESTING DUE TO MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A CRITICAL PUMP ERROR. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER STATED THAT THEY SAW A RED X ON THE DISPLAY. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642127 640G INSULIN PUMP MMT-1712K PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1712K HG22NEF 00643169999770

Patients

Seq Age Sex Outcome Treatment
1