NA DEVICE NOT SUSPECT. DEVICE IS A SELECTAIR
Report
- Report Number
- 1419507-2007-00001
- Event Type
- Death
- Date Received
- July 20, 2007
- Date of Event
- May 28, 2007
- Report Date
- July 20, 2007
- Manufacturer
- ROHO, INC.
- Product Code
- FNM
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NURSE
Narratives
THE ENTIRE SELECTAIR/TRANQUIL CARE SYSTEM WAS FUNCTIONING CORRECTLY. THE USER INSTRUCTIONS WERE NOT FOLLOWED BY THE FACILITY IN THAT THE PROPER ADJUSTMENT OF THE COMFORT LEVEL WAS NOT MADE TO THE CORRESPONDING PT'S WEIGHT AND THE PT'S BODY WAS NOT SUNKEN FORTY TO FIFTY PERCENT INTO THE MATTRESS. THE COMFORT LEVEL AS SET BY THE FACILITY AT THE TIME OF THE INCIDENT WOULD HAVE BEEN CORRECT FOR A PT WEIGHING 300 LBS NOT THE 134.5 LBS WHICH WAS RECORDED IN THE PT'S RECORDS. THIS ALLOWED THE PT'S BODY TO REST ENTIRELY ON THE SURFACE OF THE MATTRESS. THIS PERMITTED THE PT TOTALLY UNENCUMBERED MOVEMENT WHICH IS CONTRARY TO THE INSTRUCTIONS OF THE MANUFACTURER.
FROM THE FACILITY 2007. THE RESIDENT SLID OFF LEFT SIDE OF BED. ELBOW PROTRUDED THROUGH HORIZONTAL BARS OF UPPER SIDE RAIL AND HE WAS WEDGED BETWEEN THE RAIL AND MATTRESS WITH HIS WRIST AGAINST HIS THROAT. CORNER: CAUSE OF DEATH: POSITIONAL ASPHYXIA. MANNER OF DEATH: ACCIDENTAL. FROM MATTRESS PROVIDER FROM APPROX SIX WEEKS EARLIER CONVERSATION W/ THE SNF AND THREE DAYS LATER SITE VISIT: RESIDENT WAS ALERT AND ORIENTED W/ LIMITED MOBILITY. USED RAIL TO ASSIST W/ INDEPENDENT TURNING. AT 10:30PM, APPROX ONE MONTH EARLIER, RESIDENT WAS SEEN BY NURSING ASSISTANT TO BE LYING ON SIDE ASLEEP. SHIFT CHANGE ROUNDS FOUND RESIDENT 11:15-11:20PM OFF SIDE OF BED. BEDDING APPEARED TO HAVE SLID W/PT. PT KNEES ON FLOOR, BUTTOCKS ON HEELS, LEFT ELBOW PROTRUDING THROUGH SIDE RAIL OPENING, LEFT WRIST AGAINST THROAT, HEAD AGAINST MATTRESS, SHOULDER AGAINST FRAME, ENTRAPPED BETWEEN HORIZONTAL OF SIDE RAIL AND MATTRESS. RAIL 1/4-1/3 SIDE RAIL IN UP POSITION. PRELIMINARY AUTOPSY "POSITIONAL ASPHYXIATION." MATTRESS WAS SECURELY ATTACHED TO THE BED FRAME WITH VELCRO STRAPS; NO GAPS ON EITHER SIDE BETWEEN RAILS AND MATTRESS. MATTRESS, HOWEVER, WAS SET AT AN OVER-INFLATED CONDITION; FOR 300LB PERSON NOT 138LBS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NA DEVICE NOT SUSPECT. DEVICE IS A SELECTAIR | LOW AIR LOSS MATTRESS REPLACEMENT | FNM | ROHO, INC. | SA2SHORTMATT | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Death | 1/4-1/3 SIDE RAILS| RESIDENT ON A JOERNS B-50 MODEL BED W/ |