ESPIRIT VENTILATOR
Report
- Report Number
- 2031642-2007-00110
- Event Type
- Other
- Date Received
- July 20, 2007
- Date of Event
- June 28, 2007
- Report Date
- June 28, 2007
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
NA
THE CUSTOMER REPORTED, THE VENTILATOR WAS FAILING SHORT SELF TEST (SST) DUE TO SAFETY VALVE OPEN (SVO). THE UNIT BEING TESTED, THEREFORE THERE WAS NO PT INVOLVEMENT OR HARM, AND ALARMS WERE NOT APPLICABLE. THE RESPIRONICS SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE SST TEST FAILURE. THE SERVICE TECHNICIAN PERFORMED BACK PRESSURE TESTING OF THE EXPIRATORY FILTER. THE SERVICE TECHNICIAN REPORTED, THE BACKPRESSURE TEST OF THE EXPIRATORY FILTER MEASURED 87CM H20 PRESSURE. INDICATING A GROSS OCCLUSION. THE TEST SPECIFICATION IS <4CMH2O. AS SPECIFIED, A DETECTED OCCLUSION WILL CAUSE THE VENTILATOR TO ALARM AND OPEN THE SAFETY VALVE UNTIL THE OCCLUSION IS RESOLVED. THE SERVICE TECHNICIAN REPLACED THE EXPIRATORY FILTER TO CORRECT THE PROBLEM. EXTENDED SELF TESTING (EST) WAS PERFORMED AND PASSED TO OPERATING SPECIFICATIONS. USER MAINTENANCE CONTRIBUTED TO THIS EVENT DUE TO AN OCCLUDED FILTER. FILTER REPLACEMENT MUST BE PERFORMED AT MANUFACTURER RECOMMENDED INTERVALS. THE CUSTOMER WAS INFORMED OF THE SERVICE TECHNICIAN'S FINDINGS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPIRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA YR | REUSABLE EXPIRATORY FILTER |