FDA Adverse Event Other Summary report: N

ESPIRIT VENTILATOR

MDR report key: 883705 · Received July 20, 2007

Report

Report Number
2031642-2007-00110
Event Type
Other
Date Received
July 20, 2007
Date of Event
June 28, 2007
Report Date
June 28, 2007
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

NA

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THE VENTILATOR WAS FAILING SHORT SELF TEST (SST) DUE TO SAFETY VALVE OPEN (SVO). THE UNIT BEING TESTED, THEREFORE THERE WAS NO PT INVOLVEMENT OR HARM, AND ALARMS WERE NOT APPLICABLE. THE RESPIRONICS SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE SST TEST FAILURE. THE SERVICE TECHNICIAN PERFORMED BACK PRESSURE TESTING OF THE EXPIRATORY FILTER. THE SERVICE TECHNICIAN REPORTED, THE BACKPRESSURE TEST OF THE EXPIRATORY FILTER MEASURED 87CM H20 PRESSURE. INDICATING A GROSS OCCLUSION. THE TEST SPECIFICATION IS <4CMH2O. AS SPECIFIED, A DETECTED OCCLUSION WILL CAUSE THE VENTILATOR TO ALARM AND OPEN THE SAFETY VALVE UNTIL THE OCCLUSION IS RESOLVED. THE SERVICE TECHNICIAN REPLACED THE EXPIRATORY FILTER TO CORRECT THE PROBLEM. EXTENDED SELF TESTING (EST) WAS PERFORMED AND PASSED TO OPERATING SPECIFICATIONS. USER MAINTENANCE CONTRIBUTED TO THIS EVENT DUE TO AN OCCLUDED FILTER. FILTER REPLACEMENT MUST BE PERFORMED AT MANUFACTURER RECOMMENDED INTERVALS. THE CUSTOMER WAS INFORMED OF THE SERVICE TECHNICIAN'S FINDINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPIRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR REUSABLE EXPIRATORY FILTER