FDA Adverse Event Injury Summary report: N

REVERSE SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø39/+0MM

MDR report key: 8836271 · Received July 29, 2019

Report

Report Number
3005180920-2019-00644
Event Type
Injury
Date Received
July 29, 2019
Date of Event
July 8, 2019
Report Date
July 29, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706292
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 19 JULY 2019: LOT 173418: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 3.07.2017. EXPIRATION DATE: 2022-06-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ONE SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: SHOULDER DISLOCATION OCCURRED FEW MONTHS AFTER REVERSE SHOULDER ARTHROPLASTY. THESE EVENTS ARE NORMALLY ORIGINATED BY PROGRESSION OF DISEASE TO THE SOFT TISSUES OR INSUFFICIENT RE-ESTABLISHMENT OF SOFT TISSUE TENSION AFTER THE OPERATION. NO FURTHER CONCLUSION CAN BE DRAWN WITH THE ELEMENTS AT HAND. ADDITIONAL IMPLANT INVOLVED IN THE EVENT, BATCH REVIEW PERFORMED ON 19 JULY 2019. REVERSE SHOULDER SYSTEM 04.01.0003 STD HUMERAL DIAPHYSIS - CEMENTLESS - 8 (K170452) LOT. 176936. LOT 176936: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30.11.2017. EXPIRATION DATE: 2022-11-09. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY PERFORMED AFTER 4 MONTHS FROM THE PRIMARY DUE TO PAIN CAUSED BY A DISLOCATION OF THE GLENOSPHERE FROM THE LINER (THE CAUSE OF THE DISLOCATION IS UNKNOWN, NO SIGN OF INFECTION PRESENT). THE SURGEON REVISED THE STEM, LINER AND METAPHISIS SUCCESSFULLY. THE STEM SIZE HAS BEEN REDUCED FROM 8 TO 6.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629305 REVERSE SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø39/+0MM REVERSE SHOULDER HUMERAL HC LINER PHX MEDACTA INTERNATIONAL SA 173418 07630040706292

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention