FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP MMT-751LNAS

MDR report key: 8830357 · Received July 26, 2019

Report

Report Number
2032227-2019-36873
Event Type
Injury
Date Received
July 26, 2019
Date of Event
May 1, 2019
Report Date
March 9, 2020
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
00763000161170
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTIONING PROPERLY. NO BUTTON ERROR ALARM DURING TESTING. NO MOISTURE OR CORRODED KEYPAD FOUND. NO FLATTED KEYPAD. CONNECTOR WAS INSPECTED AND LOCKED ON LCD BOARD. DEVICE PASSED DISPLACEMENT TEST, REWIND TEST, BASIC OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND OCCLUSION TEST, NO ANOMALY NOTED DURING TESTING.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPOGLYCEMIA AND KEYPAD ANOMALY. CUSTOMER EXPERIENCED MULTIPLE BLOOD GLUCOSE VALUE SUCH AS 48 MG/DL, 54 MG/DL, 58 MG/DL AND 60 MG/DL. THE CUSTOMER DECLINED TROUBLESHOOTING FOR LOW BLOOD GLUCOSE LEVEL. THE CUSTOMER WAS TREATED WITH ALMOND ILK AND PEANUT BUTTER. CUSTOMER STATED THAT THE BUTTONS ON THE INSULIN PUMP ARE STARTING TO STICK AND UNEXPECTEDLY GOES LOW. CUSTOMER STATED THAT THE INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARM. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
625730 530G INSULIN PUMP MMT-751LNAS ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751LNAS A6751LNASJ 00763000161170

Patients

Seq Age Sex Outcome Treatment
1 31 YR Other FRN-MMT-332-RSVR, UNOMED INF SET