FDA Adverse Event
Malfunction
Summary report: N
HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT
MDR report key: 8830086
·
Received July 26, 2019
Report
- Report Number
- 3005180920-2019-00612
- Event Type
- Malfunction
- Date Received
- July 26, 2019
- Date of Event
- June 26, 2019
- Report Date
- July 26, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- UDI-DI
- 07630040742115
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
BATCH REVIEW PERFORMED ON 09 JULY 2019: LOT 1854515: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, THREE OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THIS SAME LOT. VISUAL INSPECTION PERFORMED BY R&D HIP PROJECT MANAGER: THE SPRING BALL PLUNGER IS CAME APART AS DESCRIBED IN THE COMPLAINT. IT IS POSSIBLE THIS OCCURRENCE WAS INFLUENCED BY VARIATION IN PRODUCTION/ASSEMBLY HOWEVER THIS CANNOT BE CONFIRMED. EVALUATION OF THE SOLUTION ON GOING.
Description of Event or Problem · 1
DURING SURGERY THE LOCKING BALL FELL OUT OF THE HANDLE. THE BALL IS NOT FALLING IN THE PATIENT. THE SURGERY HAS BEEN COMPLETED USING ANOTHER INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 623984 | HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT | SURGICAL INSTRUMENT FOR HIP SURGERY | LXH | MEDACTA INTERNATIONAL SA | 1854515 | 07630040742115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |