FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE

MDR report key: 8828121 · Received July 25, 2019

Report

Report Number
9610847-2019-00481
Event Type
Malfunction
Date Received
July 25, 2019
Date of Event
July 10, 2019
Report Date
September 5, 2019
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FPA
UDI-DI
30382903851004
PMA / PMN Number
K013621
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9007914. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, VISUAL EVALUATION OF THE SUBMITTED SAMPLES AND THE RESULTING REVIEW OF THE MANUFACTURING PROCESS DETERMINED THAT THE MATERIAL IS EXCESS ADHESIVE THAT IS USED IN THE PACKAGING PROCESS. THIS DEFECT IS RELATED WITH LUMPS OF ADHESIVE ACCUMULATED IN THE ROLLERS.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND IN THE BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE PACKAGING BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THERE IS FOREIGN MATERIAL IN THE PACKAGE OF Q-SYTE."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND IN THE BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE PACKAGING BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THERE IS FOREIGN MATERIAL IN THE PACKAGE OF Q-SYTE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
623214 BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE INTRAVASCULAR ADMINISTRATION SET FPA BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 9007914 30382903851004

Patients

Seq Age Sex Outcome Treatment
1 Other