PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAL
Report
- Report Number
- 2032227-2019-35987
- Event Type
- Malfunction
- Date Received
- July 25, 2019
- Date of Event
- July 15, 2019
- Report Date
- August 8, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169513129
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
DEVICE RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE AT ELECTRONIC ASSEMBLY. ALSO, MOISTURE DAMAGE MOTOR DURING VISUAL INSPECTION. UNABLE TO PERFORM OPERATING CURRENTS, SELF TEST, REWIND TEST AND DISPLACEMENT TEST DUE TO BLANK DISPLAY.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD BLANK DISPLAY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 252 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DISPLAY DID NOT RETURNED BACK. CUSTOMER STATED THAT THE CONTACTS ON THE BATTERY CAP WERE NOT MISSING OR DAMAGED, BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED AND SPRING IS NEITHER DAMAGED NOR CORRODED. THE CUSTOMER WAS ADVISED TO REPLACE AND TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619043 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAL | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723RNAL | A1723RNALJ | 00643169513129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR |