FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 8824600 · Received July 24, 2019

Report

Report Number
3004753838-2019-055791
Event Type
Malfunction
Date Received
July 24, 2019
Date of Event
July 7, 2019
Report Date
September 10, 2019
Manufacturer
DEXCOM, INC.
Product Code
PQF
UDI-DI
10386270000221
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY.

Description of Event or Problem · 1

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616742 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQF DEXCOM, INC. 9438-06 7254305 10386270000221

Patients

Seq Age Sex Outcome Treatment
1 46 YR