VITROS 4600 CHEMISTRY SYSTEM
Report
- Report Number
- 1319681-2019-00064
- Event Type
- Malfunction
- Date Received
- July 24, 2019
- Date of Event
- June 27, 2019
- Report Date
- July 24, 2019
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT A LOWER THAN EXPECTED VITROS VALPROIC ACID (VALP) RESULT WAS OBTAINED FROM A VITROS TDM PERFORMANCE VERIFIER (PV) CONTROL FLUID WHEN RUN ON A VITROS 4600 CHEMISTRY SYSTEM. THE MOST LIKELY ASSIGNABLE CAUSE OF THE LOWER THAN EXPECTED VITROS VALP RESULT WAS AN INSTRUMENT EVENT RELATED ISSUE. AN UNACCEPTABLE WITHIN RUN VITROS VALP PRECISION TEST INDICATED THE VITROS 4600 SYSTEM WAS NOT PERFORMING AS EXPECTED. AN ACCEPTABLE WITHIN RUN VITROS VALP PRECISION TEST WAS OBTAINED AFTER SERVICE ACTIONS PERFORMED BY AN ORTHO FIELD ENGINEER WHICH INCLUDED CHANGING THE MICROIMMUNOASSAY (UIA) PROBOSCIS, CHANGING THE PHOTOMETER LAMP, PERFORMING A WATER BLANK, AND PERFORMING ALL CUVETTE INCUBATOR ADJUSTMENTS. THEREFORE, A VITROS 4600 SYSTEM ISSUE IS THE LIKELY CAUSE OF THE EVENT. THERE IS NO EVIDENCE THAT THE EVENT WAS CAUSED BY A REAGENT ISSUE WITH VITROS VALP LOT 2511-27-7325. IN ADDITION, CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD POINT TO A POTENTIAL SYSTEMIC ISSUE WITH VITROS VALP, LOT 2511-27-7325.
A CUSTOMER OBTAINED A LOWER THAN EXPECTED VITROS VALPROIC ACID (VALP) RESULT FROM A VITROS TDM PERFORMANCE VERIFIER (PV) CONTROL FLUID WHEN RUN ON A VITROS 4600 CHEMISTRY SYSTEM. VITROS TDM PVI VALP RESULT OF 16.8 UG/ML VERSUS THE MIDPOINT OF THE ROM (RANGE OF MEAN) 28.6 UG/ML. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF UNDETECTED. THE LOWER THAN EXPECTED VITROS VALP RESULT WAS OBTAINED FROM A NON-PATIENT FLUID. HOWEVER THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WERE NOT AFFECTED OR WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. ORTHO WAS NOT MADE AWARE OF ANY ALLEGATION OF ACTUAL PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO).COMPLAINT NUMBERS: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 617086 | VITROS 4600 CHEMISTRY SYSTEM | CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |