INRATIO
Report
- Report Number
- 2954730-2007-00330
- Event Type
- Malfunction
- Date Received
- July 16, 2007
- Date of Event
- July 13, 2007
- Report Date
- July 13, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
TS UPDATED THE CASE ON 7/13/2007. DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE 2007- INRATIO 5.2, 3.7, 1.4; LAB 4.5, 2.2, 3.1; MEAN 4.9, 3.0, 2.25; CONFIDENCE LIMITS 2.8 - 7.2, 1.8 - 4.2, 1.4 - 3.1. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.
TS UPDATED CASE ON 7/13/2007: CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE 2007; INRATIO 5.2 - 3.7 AND 1.4; LAB 4.5 - 2.2 AND 3.1.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |