FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 8821750
·
Received July 24, 2019
Report
- Report Number
- 2518422-2019-01611
- Event Type
- Malfunction
- Date Received
- July 24, 2019
- Date of Event
- July 9, 2019
- Report Date
- July 9, 2019
- Manufacturer
- PHILIPS RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO A THIRD PARTY SERVICE CENTER FOR EVALUATION. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT A THIRD PARTY SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR WAS REPLACED TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 616043 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | PHILIPS RESPIRONICS INC. | 1054260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |