FDA Adverse Event Injury Summary report: N

EVOLUT R TRANSCATHETER AORTIC VALVE

MDR report key: 8818877 · Received July 23, 2019

Report

Report Number
2025587-2019-02308
Event Type
Injury
Date Received
July 23, 2019
Date of Event
July 1, 2019
Report Date
July 23, 2019
Manufacturer
MEDTRONIC HEART VALVES DIVISION
Product Code
NPT
PMA / PMN Number
P130021
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CITATION: LEE J ET AL. EARLY TRANSCATHETER AORTIC VALVE FAILURE ACCOMPANIED WITH LEAFLET PERFORATION. KOREAN CIRC J. 2019 JUL;49(7): 642-643. HTTPS://DOI.ORG/10.4070/KCJ.2018.0457. EARLIEST DATE OF PUBLISH USED FOR EVENT DATE. NO UNIQUE DEVICE IDENTIFIER (SERIAL/LOT) NUMBERS WERE PROVIDED; WITHOUT THIS INFORMATION, IT COULD NOT BE DETERMINED WHETHER THESE OBSERVATIONS HAVE BEEN PREVIOUSLY REPORTED. WITHOUT RETURN OF THE PRODUCT NO DEFINITIVE CONCLUSION CAN BE MADE REGARDING THE CLINICAL OBSERVATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION VIA LITERATURE REGARDING A (B)(6) YEAR-OLD FEMALE PATIENT WHO PRESENTED WITH AGGRAVATED DYSPNEA AND CLASS IV HEART FAILURE. THE PATIENT HAD BEEN IMPLANTED WITH A 26 MM MEDTRONIC EVOLUT-R BIOPROSTHETIC VALVE 17 MONTHS PRIOR (NO SERIAL NUMBER PROVIDED). THE PATIENT WAS LATER DIAGNOSED WITH METHICILLIN-SENSITIVE STAPHYLOCOCCUS EPIDERMIDIS, ACUTE PYELONEPHRITIS AND RIGHT KNEE SEPTIC ARTHRITIS REQUIRING SURGICAL DEBRIDEMENT. A TRANSTHORACIC ECHOCARDIOGRAM (TTE) AND TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) REVEALED SEVERE TRANSVALVULAR REGURGITATION. DURING SURGICAL INTERVENTION THE BIOPROSTHETIC VALVE RIGHT CORONARY CUSP WAS NOTED AS PERFORATED, AND THE VALVE WAS EXPLANTED. POST-EXPLANT HISTOPATHOLOGICAL ANALYSIS REVEALED EVIDENCE OF ENDOCARDITIS AND STRUCTURAL VALVE DETERIORATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS OR PRODUCT PERFORMANCE ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611614 EVOLUT R TRANSCATHETER AORTIC VALVE AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT MEDTRONIC HEART VALVES DIVISION EVOLUTR-23

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R